19
P/N VT-108723-US, Manual revision: C
Vante
®
SEBRA
®
MINI
™
Tube Sealer Power Source
Battery charging
Caution: DO NOT attempt to charge non-chargeable batteries. Damage to
equipment could occur.
Caution: The battery charger should never be used as an AC adapter by
plugging the RF power source connector into the battery charger connector.
The Power Source will not function properly and the charger may be damaged.
1. Plug the battery charger into an AC power source (wall outlet) and verify
that the green power "ON" indicator light is illuminated.
Note: Check to see that the power cord plug matches the power receptacle for
the country in which the battery charger is being used. If it does not, or if the
green power "ON" indicator light does not illuminate, contact Vante
Customer
®
Care Center.
2. Once the battery pack is removed from the RF power source (refer to
battery replacement page18), plug its red and black connector into the
matching connector on the battery charger.
Caution: The battery pack and charger connectors are keyed to properly
engage. Be sure that the RED battery pack connector is engaged into the RED
battery charger connector, and the BLACK into the BLACK. If they do not easily
engage, do not attempt to force the connectors together. Improper battery
connection will result in damage to the battery pack and/or charger and will void
any warranties.
Caution: Charge the battery pack only with the Vante
®
SEBRA
®
Model 2389
Rapid Charger to ensure safety and prevent battery damage.
Note: If amber charge status light fails to illuminate after connection, check to
see that the battery pack is properly connected to the charger. If green power
"ON" indicator light flashes, leave battery pack on charger for one hour. If
amber charge status light still fails to illuminate after one hour, disconnect
battery pack from charger and contact Vante Customer Care Center for
battery pack replacement.
3. Leave battery pack on battery charger until completely charged. Battery
pack is completely charged when amber charge status indicator light
flashes every 1/8-second.
Charge status is represented by the amber charge status indicator light as
Table 4-1, “Charge Indicator Light Status,” on page 21.
Summary of Contents for SEBRA MINI 2380
Page 2: ...ii 2007 2011 2017 Vante Inc All rights reserved...
Page 4: ...Vante SEBRA MINI Tube Sealer Power Source P N VT 108723 US Manual revision C iv...
Page 30: ...P N VT 108723 US Manual revision C Vante SEBRA MINI Tube Sealer Power Source 24...
Page 32: ...Vante Inc 3480 E Britannia Dr Suite 120 Tucson AZ 85706 USA www vante com...