background image

Turn the coarse/fine focus controls (figure 14a) to adjust the image focus. The larger, 

outer knob is used for coarse adjustment, the smaller knob for fine adjustment.
Turning the coarse/fine focus controls raises and lowers the stage vertically. One 

complete turn of the fine focusing knob raises or lowers the stage 0.3mm; the smallest 

graduation refers to 2 microns of vertical movement. One complete turn of the coarse 

focusing knob raises or lowers the stage 3.6mm. To ensure long life, turn the focusing 

knobs slowly and uniformly.
The focusing tension control is located just inside of the right-hand focus control knob 

(figure 14b). For tighter tension, use the included tension control wrench to turn the 

control in a clockwise motion. For looser tension, turn the control in a counterclockwise 

motion.

Illuminator Focusing: The condenser can be raised and lowered with the substage 

adjustment knob (figure 15a) to focus the light for optimal illumination.
Aperture Iris Diaphragm Adjustment: In order to achieve the best contrast, resolution, 

and depth of field, the aperture diaphragm should be set to approximately 70-80% of the 

field to match the objective’s numerical aperture. Adjust the aperture iris diaphragm with 

the aperture iris diaphragm lever (figure 16a) located at the bottom of the condenser.
Field Iris Diaphragm Adjustment: The field diaphragm (figure 16b) is opened or closed 

by rotating the adjustment ring and can be used to limit unwanted peripheral light from 

entering the condenser and interfering with the image quality.
The mechanical stage X-Y controls, located underneath the right-hand side of the stage 

(figure 14c), provide easy and accurate positioning of the sample. One complete turn 

of the longitudinal (X) control (lower half of the stage controls) will move the specimen 

34mm left or right. One complete turn of the transverse (Y) control (upper half of the 

stage controls) will move the specimen 20mm front or back.
The spring-loaded slide holder can be removed for users who prefer to not use it . 

Simply unthread the slide holder knobs (figure 16c), which lock the slide holder on 

the stage, and remove the slide holder.

Variable Lighting Control and Power Switch

1

7

6

5

4

3

Figure 14

Figure 15

Focusing and Mechanical Stage Mechanisms

The rocker switch located on the right hand side of the microscope (figure 12) turns 

the power on or off.  To turn the power on press the rocker on the end marked “|”.  To 

turn the power off press the rocker on the end marked “O”.

10

Operation

All Models

Figure 12

Figure 13

All Models

8

VanGuard 1300 Series microscopes are equipped with a variable lighting control, 

located on the left side of the microscope, which allows the user to set the lighting 

anywhere between off and full brightness (figure 13). To increase the brightness roll 

the wheel towards the front of the microscope.

1

Figure 16

[a]

[b]

[c]

[a]

[c]

[a]

[b]

Summary of Contents for 1320BR

Page 1: ...Operation Manual Covering Models 1320BR 1321BRi 1323PHi 1330BR 1331BRi 1333PHi 1300 Series ...

Page 2: ...N A Abbe condenser Phase contrast models come with a 1 25 N A Zernike condenser with phase annulus rings for 10X 20X 40X and 100X also has a 0 setting for brightfield work All condensers are mounted on a rack and pinion focusing mechanism and feature spring loaded centering knobs and iris diaphragm Illumination 20W 6V variable quartz halogen light source Comes with blue green phase contrast only a...

Page 3: ...chromatic Objective 1 ea 100X Achromatic Objective 1 ea Dispersion Filter 1 ea Blue Filter 1 ea Power Cable 1 ea Spare Halogen Lamp 1 ea Lamp Replacement Tool 1 ea Spare Fuse 1 ea Dust Cover 1 ea Operation Manual 1 ea Model 1331BRi Trinocular Head 1 ea Stand 1 ea Brightfield Condenser 1 ea 10X Eyepiece 2 ea 4X Infinity Corrected Plan Objective 1 ea 10X Infinity Corrected Plan Objective 1 ea 40X In...

Page 4: ...r Condenser Set Screw Drop In Filter Holder Brightfield Abbe Condenser Brightfield models only Interpupillary Scale Up Stop Mechanism Model 1331BRi Shown Brightfield Interpupillary Adjustment Mechanism Collector Assembly Trinocular Port Trinocular models only Trinocular Port Dust Cover Trinocular models only 4 ...

Page 5: ...hase Contrast Condenser Phase models only Fine Focus Control Base Stand Mechanical Stage Slide Holder Slide Holder Knobs Focusing Tension Control Knob Coarse Focus Control Prism Slider Trinocular models only Model 1333PHi Shown Phase Contrast Kohler Iris Diaphragm Centering Screws Power Switch Kohler Iris Diaphragm Phase Annulus Turret 5 ...

Page 6: ...withAdapters and Microscopy Software 10 Megapixels Cleaning Supplies Consumables Part Number Description 45000 OT001 VEE GEE Optical Tissue 1200 IOG Immersion Oil Low Viscosity 1 4 fl oz OtherAccessories Part Number Description 1400 MCL Hard Shell Vinyl Microscope Case 1200 PAK 38 Polarizer Analyzer Kit Parts Accessories For information about parts accessories or service contact your dealer direct...

Page 7: ...ulling upwards and out at the same time The collector assembly is held in place by magnets and will come loose with a minimal amount of force Thelamp figure4b shouldbecenteredinthelampsocket figure4c Ifitisnot graspthe lamp using a tissue being careful not to touch the glass envelope with your bare fingers and reposition the lamp so that the filament is positioned at the center of the socket Reins...

Page 8: ... the substage assembly fully by turning the substage adjustment knob counter clockwise Be careful not to crash the condenser lens into the objectives OpentheKohlerfieldirisdiaphragmtothelargestsettingbyrotatingtheadjustment ring figure 5a clockwise Slide the aperture diaphragm lever figure 6a to the right to close the aperture iris Closing the aperture diaphragm will reduce the field so that a sma...

Page 9: ...centering telescope should resemble rings superimposed on one another figures 9 10 1 2 3 4 5 6 7 8 9 10 Phase Contrast Models Figure 9 Figure 10 Figure 11 Insertasmallslottedscrewdriverintooneofthetwoaccessslotsintherearofthecondenser figure11 toaccesstheadjustmentscrews orpreferablyusetwoscrewdriversatthesame time one in each access slot While looking through the centering telescope watch the whi...

Page 10: ...fielddiaphragm figure16b isopenedorclosed byrotatingtheadjustmentringandcanbeusedtolimitunwantedperipherallightfrom entering the condenser and interfering with the image quality ThemechanicalstageX Ycontrols locatedunderneaththeright handsideofthestage figure 14c provide easy and accurate positioning of the sample One complete turn of the longitudinal X control lower half of the stage controls wil...

Page 11: ...terpupillary adjustment the distance between the eyepieces is made through a folding action The Seidentopf design allows for a folding adjustment which is quickly and easily done for each user figure 17 Diopter adjustment allows for proper optical correction based on each individual s eyesight This adjustment is easily made and is recommended prior to each use by different users to prevent eyestra...

Page 12: ...ith this microscope Remove the camera port dust cap figure 21b then slide the adapter into the camera port Pull the prism slider figure 21c completely out to divert the image to the camera port The 1300 Series trinocular microscopes utilize an 80 20 split sliding prism This split prism diverts 80 of the light to the camera port and the remaining 20 to the eyepieces when engaged This allows the eye...

Page 13: ...he objective Note Under no circumstances should an oil immersion objective be left sitting in oil for an extended period of time Exceptionally long immersion periods can cause oil to penetrate the objective s sealant and obscure the optics which is not covered under warranty Note You should avoid getting oil on any objective other than the 100X as they are not sealed Pay particular attention to th...

Page 14: ...6V Halogen Cat No 1400 20WHL Figure 23 a b a c Figure 24 Figure 25 Whenreplacing firstinsertthenewlampintotheincludedlampreplacementtool figure 24 Then use the lamp replacement tool to insert the lamp into the lamp socket figure 25 Thelampshouldbeproperlycenteredwithinthelampsocketforoptimalillumination Remove the lamp replacement tool after insertion Reinstall the collector assembly making sure t...

Page 15: ...n the bonding material holding the lenses should never be used as a cleaner Periodically your VanGuard Microscope should be fully serviced by a qualified service technician In order to keep dust and debris out of the optical pathways always keep the camera port and eyetubes covered with either eyepieces or dust caps and always use the dust cover when the microscope is not in use Caring for your Va...

Page 16: ...Symptom Image through eyepiece s too dim Step Possible Cause How To Test Test Result Solution For More Info 1 5 4 3 2 Variable lighting control VLC is turned down too far Visually inspect VLC is turned up VLC is turned down Go to next step Turn VLC up Page 10 Light bulb not centered properly Visually inspect Bulb is centered Bulb is not centered Go to next step Center the bulb Condenser not center...

Page 17: ...copes Page 18 Page 13 Symptom The stage keeps drifting downward Step Possible Cause How To Test Test Result Solution For More Info 1 Focus tension is set too loose Check focus tension control Tension too loose Tension set correctly Increase tension Contact dealer or VanGuard Microscopes Page 18 Page 10 Symptom No image visable in eyepiece s or trinocular port Step Possible Cause How To Test Test R...

Page 18: ...air or replacement of parts by the End User or an unauthorized third party while under warranty 4 Consumable items such as but not limited to bulbs have failed This warranty expressly excludes transportation damage and adjustment or readjustment In no case shall VEE GEE Scientific be liable to the Buyer or any person for any special indirect incidental or consequential damage whether claims are ba...

Page 19: ...bjectives 10X 0 25 NA 7 4mm WD Plan Achromatic 20X 0 40 NA 5 8mm WD 40X 0 65 NA 0 70mm WD 100X oil 1 25 NA 0 25mm WD Stage Dimensions 135mm x 140mm Stage Motion Right Hand Coaxial Control Rack Pinion Drive Stage Movement Range 50 x 80mm Focusing Movement Coaxial Coarse Fine Controls Safety Up Stop Focusing Range 40mm Focusing Graduation 2 Microns Division Brightfield Condenser 1 25 NA Abbe Condens...

Page 20: ...Printed in the USA VGMNL070219 1300BRIPHI veegee com 800 423 8842 sales veegee com techsupport veegee com veegee com pages technical support service ...

Reviews: