background image

Setup

8

Centering

:  Although the condenser comes pre-centered you may find it necessary to 

check it occasionally. A simple method for centering is as follows:  

1

2

3

All Models

Figure 5

Figure 6

4

Figure 8

70-80%

Figure 7a

Figure 7b

Figure 7c

Figure 7d

Vertical Focusing

:  The condenser can be raised and lowered with the substage 

adjustment knob to focus the light for optimal illumination.

Aperture Iris Diaphragm Adjustment

:  To achieve the best contrast, resolution, and 

depth of field, the aperture diaphragm should be set to approximately 70-80% of the field 

(figure 8) to match the objective’s numerical aperture. Adjust the aperture iris diaphragm 

with the aperture iris diaphragm lever (figure 6a) located at the bottom of the condenser.

Field Iris Diaphragm Adjustment

:  The field diaphragm can be used to limit unwanted 

peripheral light from entering the condenser and interfering with the image quality.

[a]

Substage Adjustments

•  Rotate the nosepiece until the 10X objective is in the light path. For phase contrast 

models set the phase annulus turret to the “0” position.

•  Raise the substage assembly fully by turning the substage adjustment knob   

counter-clockwise. Be careful not to crash the condenser lens into the objectives.

• 

Open the Kohler field iris diaphragm to the largest setting by rotating the adjustment 

ring (figure 5a) clockwise.

• 

Slide the aperture diaphragm lever (figure 6a) to the right to close the aperture iris.

• 

Closing the aperture diaphragm will reduce the field so that a small white hexagon 

is visible within a black field (figure 7a). Remove one of the eyepieces and view 

the image directly through the empty eyetube. Focus the hexagon by adjusting the 

substage adjustment knob. This white hexagon should be centered in the black 

field.  If not, using a 1.5mm hex wrench first loosen the two condenser centering 

screws located at the rear of the condenser mount, then reposition the condenser 

until the iris is properly centered within the black field (figure 7b). Finally, tighten the 

condenser centering screws back down.

• 

Fine tune by opening the aperture iris diaphragm until the white hexagon almost fills 

the entire field (figure 7c), and then readjust (figure 7d) by using the instructions in 

the previous step. After centering the condenser, open the aperture iris diaphragm 

slightly wider than the field of view.

[a]

Summary of Contents for 1320BR

Page 1: ...Operation Manual Covering Models 1320BR 1321BRi 1323PHi 1330BR 1331BRi 1333PHi 1300 Series ...

Page 2: ...N A Abbe condenser Phase contrast models come with a 1 25 N A Zernike condenser with phase annulus rings for 10X 20X 40X and 100X also has a 0 setting for brightfield work All condensers are mounted on a rack and pinion focusing mechanism and feature spring loaded centering knobs and iris diaphragm Illumination 20W 6V variable quartz halogen light source Comes with blue green phase contrast only a...

Page 3: ...chromatic Objective 1 ea 100X Achromatic Objective 1 ea Dispersion Filter 1 ea Blue Filter 1 ea Power Cable 1 ea Spare Halogen Lamp 1 ea Lamp Replacement Tool 1 ea Spare Fuse 1 ea Dust Cover 1 ea Operation Manual 1 ea Model 1331BRi Trinocular Head 1 ea Stand 1 ea Brightfield Condenser 1 ea 10X Eyepiece 2 ea 4X Infinity Corrected Plan Objective 1 ea 10X Infinity Corrected Plan Objective 1 ea 40X In...

Page 4: ...r Condenser Set Screw Drop In Filter Holder Brightfield Abbe Condenser Brightfield models only Interpupillary Scale Up Stop Mechanism Model 1331BRi Shown Brightfield Interpupillary Adjustment Mechanism Collector Assembly Trinocular Port Trinocular models only Trinocular Port Dust Cover Trinocular models only 4 ...

Page 5: ...hase Contrast Condenser Phase models only Fine Focus Control Base Stand Mechanical Stage Slide Holder Slide Holder Knobs Focusing Tension Control Knob Coarse Focus Control Prism Slider Trinocular models only Model 1333PHi Shown Phase Contrast Kohler Iris Diaphragm Centering Screws Power Switch Kohler Iris Diaphragm Phase Annulus Turret 5 ...

Page 6: ...withAdapters and Microscopy Software 10 Megapixels Cleaning Supplies Consumables Part Number Description 45000 OT001 VEE GEE Optical Tissue 1200 IOG Immersion Oil Low Viscosity 1 4 fl oz OtherAccessories Part Number Description 1400 MCL Hard Shell Vinyl Microscope Case 1200 PAK 38 Polarizer Analyzer Kit Parts Accessories For information about parts accessories or service contact your dealer direct...

Page 7: ...ulling upwards and out at the same time The collector assembly is held in place by magnets and will come loose with a minimal amount of force Thelamp figure4b shouldbecenteredinthelampsocket figure4c Ifitisnot graspthe lamp using a tissue being careful not to touch the glass envelope with your bare fingers and reposition the lamp so that the filament is positioned at the center of the socket Reins...

Page 8: ... the substage assembly fully by turning the substage adjustment knob counter clockwise Be careful not to crash the condenser lens into the objectives OpentheKohlerfieldirisdiaphragmtothelargestsettingbyrotatingtheadjustment ring figure 5a clockwise Slide the aperture diaphragm lever figure 6a to the right to close the aperture iris Closing the aperture diaphragm will reduce the field so that a sma...

Page 9: ...centering telescope should resemble rings superimposed on one another figures 9 10 1 2 3 4 5 6 7 8 9 10 Phase Contrast Models Figure 9 Figure 10 Figure 11 Insertasmallslottedscrewdriverintooneofthetwoaccessslotsintherearofthecondenser figure11 toaccesstheadjustmentscrews orpreferablyusetwoscrewdriversatthesame time one in each access slot While looking through the centering telescope watch the whi...

Page 10: ...fielddiaphragm figure16b isopenedorclosed byrotatingtheadjustmentringandcanbeusedtolimitunwantedperipherallightfrom entering the condenser and interfering with the image quality ThemechanicalstageX Ycontrols locatedunderneaththeright handsideofthestage figure 14c provide easy and accurate positioning of the sample One complete turn of the longitudinal X control lower half of the stage controls wil...

Page 11: ...terpupillary adjustment the distance between the eyepieces is made through a folding action The Seidentopf design allows for a folding adjustment which is quickly and easily done for each user figure 17 Diopter adjustment allows for proper optical correction based on each individual s eyesight This adjustment is easily made and is recommended prior to each use by different users to prevent eyestra...

Page 12: ...ith this microscope Remove the camera port dust cap figure 21b then slide the adapter into the camera port Pull the prism slider figure 21c completely out to divert the image to the camera port The 1300 Series trinocular microscopes utilize an 80 20 split sliding prism This split prism diverts 80 of the light to the camera port and the remaining 20 to the eyepieces when engaged This allows the eye...

Page 13: ...he objective Note Under no circumstances should an oil immersion objective be left sitting in oil for an extended period of time Exceptionally long immersion periods can cause oil to penetrate the objective s sealant and obscure the optics which is not covered under warranty Note You should avoid getting oil on any objective other than the 100X as they are not sealed Pay particular attention to th...

Page 14: ...6V Halogen Cat No 1400 20WHL Figure 23 a b a c Figure 24 Figure 25 Whenreplacing firstinsertthenewlampintotheincludedlampreplacementtool figure 24 Then use the lamp replacement tool to insert the lamp into the lamp socket figure 25 Thelampshouldbeproperlycenteredwithinthelampsocketforoptimalillumination Remove the lamp replacement tool after insertion Reinstall the collector assembly making sure t...

Page 15: ...n the bonding material holding the lenses should never be used as a cleaner Periodically your VanGuard Microscope should be fully serviced by a qualified service technician In order to keep dust and debris out of the optical pathways always keep the camera port and eyetubes covered with either eyepieces or dust caps and always use the dust cover when the microscope is not in use Caring for your Va...

Page 16: ...Symptom Image through eyepiece s too dim Step Possible Cause How To Test Test Result Solution For More Info 1 5 4 3 2 Variable lighting control VLC is turned down too far Visually inspect VLC is turned up VLC is turned down Go to next step Turn VLC up Page 10 Light bulb not centered properly Visually inspect Bulb is centered Bulb is not centered Go to next step Center the bulb Condenser not center...

Page 17: ...copes Page 18 Page 13 Symptom The stage keeps drifting downward Step Possible Cause How To Test Test Result Solution For More Info 1 Focus tension is set too loose Check focus tension control Tension too loose Tension set correctly Increase tension Contact dealer or VanGuard Microscopes Page 18 Page 10 Symptom No image visable in eyepiece s or trinocular port Step Possible Cause How To Test Test R...

Page 18: ...air or replacement of parts by the End User or an unauthorized third party while under warranty 4 Consumable items such as but not limited to bulbs have failed This warranty expressly excludes transportation damage and adjustment or readjustment In no case shall VEE GEE Scientific be liable to the Buyer or any person for any special indirect incidental or consequential damage whether claims are ba...

Page 19: ...bjectives 10X 0 25 NA 7 4mm WD Plan Achromatic 20X 0 40 NA 5 8mm WD 40X 0 65 NA 0 70mm WD 100X oil 1 25 NA 0 25mm WD Stage Dimensions 135mm x 140mm Stage Motion Right Hand Coaxial Control Rack Pinion Drive Stage Movement Range 50 x 80mm Focusing Movement Coaxial Coarse Fine Controls Safety Up Stop Focusing Range 40mm Focusing Graduation 2 Microns Division Brightfield Condenser 1 25 NA Abbe Condens...

Page 20: ...Printed in the USA VGMNL070219 1300BRIPHI veegee com 800 423 8842 sales veegee com techsupport veegee com veegee com pages technical support service ...

Reviews: