Maintenance

The eyepieces and objectives on VanGuard Microscopes are coated. They should never be wiped while dry 

as 

any dirt or dust will scratch the coating. The surfaces should either be blown off with an air canister, or 

blown 

and cleaned with an air-bulb and camel-hair brush. It is recommended to then use a lens cleaner. 

Never use anything other than lens cleaner on any optical component.

 Apply with a cotton swab for a 

minimum of wetting, then wipe the surface clean with a quality lens tissue. 

The same care instructions apply to all optical parts on this VanGuard Microscope, including the substage 

collector and the condenser lenses.
All other parts can be cleaned with a paper towel and rubbing alcohol.  Be aware that rubbing alcohol can 

break down lubricants, so be careful when cleaning near the following parts:

Stage rack and pinion gears

Focus controls

Nosepiece

Substage gears

Substage controls

Xylene, since it breaks down the bonding material holding the lenses, should never be used as a cleaner.  

Periodically your VanGuard Microscope should be fully serviced by a qualified service technician.  
In order to keep dust and debris out of the optical pathways, always keep the camera port and eyetubes 

covered (with either eyepieces or dust caps), and always use the dust cover when the microscope is not in use.

Caring for your 1200CM Microscope

1

5

4

3

6

All Models

www.calcert.com

[email protected]

1.888.610.7664

0

5

10

15

20

25

30

Summary of Contents for 1210CM, 1211CM, 1220CM, 1221CM, 1222CM, 1223CM, 1230CM, 1231CM, 1232CM,...

Page 1: ...Covering Models 1210CM 1211CM 1220CM 1221CM 1222CM 1223CM 1230CM 1231CM 1232CM 1233CM Operation Manual 1200CM Series www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30...

Page 2: ...be condenser Phase contrast models come with a 1 25 N A Zernike condenser with phase annulus rings for 10X 25X 40X and 100X also has a 0 setting for brightfield work All condensers are mounted on a ra...

Page 3: ...lan Phase Objective 1 ea 40X Plan Phase Objective 1 ea 100X Plan Phase Objective 1 ea Dispersion Filter 1 ea Blue Filter 1 ea Green Filter 1 ea Model 1230CM Trinocular Head 1 ea Stand 1 ea Brightfield...

Page 4: ...1400 CDPC 3 USB Digital Camera System 3 Megapixels for 1210 and 1220 series 1200 CDPC 3 USB DIgital Camera System 3 Megapixels for 1230 series Other Accessories Part Number Description Call for Part...

Page 5: ...ature Iris Diaphragm Adjustment Lever Fine Focus Control Substage Collector Base Stand Mechanical Stage Slide Holder Slide Holder Knobs Focusing Tension Control Knob Coarse Focus Control Camera Port T...

Page 6: ...se Annuli Centering Knobs Phase Contrast Condenser Iris Diaphragm Adjustment Light Bulb Centering Mechanism Substage Collector Set Screw 6 Light Bulb Access Door Lock Knob Swing out Filter Holder Mode...

Page 7: ...ing the objectives from their storage containers individually install each one into the nosepiece by threading them clockwise into the threaded holes of the nosepiece see figure 3 It is generally cons...

Page 8: ...small white hexagon is visible within a black field see Figure 6A Focusing of the hexagon is performed by turning the coarse fine focus controls This white hexagon is the light which is passing throu...

Page 9: ...of light are centered upon one another see Figure 8 Once the phase rings are centered remove the phase contrast centering telescope and replace with the eyepiece The phase rings are now centered for t...

Page 10: ...not to adjust light intensity The mechanical stage X Y controls located underneath the right hand side of the stage see figure 9 provide easy and accurate positioning of the sample One complete turn...

Page 11: ...er Adjustment 1 2 3 Interpupillary adjustment the distance between eyepieces is made through a folding action The Seidentopf design allows for a folding adjustment which is quickly and easily done for...

Page 12: ...mly secured with the camera port set screw Tighten the camera port set screw Pull the prism slider completely out to divert the image to the camera port The prism slider is the silver knob on the righ...

Page 13: ...omes with this microscope must be used with immersion oil in order to maintain image quality After use the objective tip needs to be wiped clean so that no oil residue remains Procedure for using imme...

Page 14: ...ld not have to force the bulb Close the light bulb access door and screw the knob back in to lock Return the microscope to normal operating position If necessary recenter the illuminator following the...

Page 15: ...the condenser lenses All other parts can be cleaned with a paper towel and rubbing alcohol Be aware that rubbing alcohol can break down lubricants so be careful when cleaning near the following parts...

Page 16: ...ble Cause How to Check Result Solution For More Info 1 Variable lighting control VLC is turned down too far Visually inspect VLC turned up Go to next step VLC turned down Turn VLC up Page 10 2 Bulb no...

Page 17: ...e Cause How to Check Result Solution For More Info 1 Up stop set too low Visually inspect Set too low Reset up stop to appropriate height Page 11 Not set too low Go to next step 2 Using the 100X objec...

Page 18: ...c or Plan Achromatic 20X 0 40 NA 0 50mm WD 0 9mm FOV 40X 0 65 NA 0 40mm WD 0 45mm FOV 100X oil 1 25 NA 0 25mm WD 0 18mm FOV Stage Dimensions 125mm x 135mm Stage Motion Right Hand Coaxial Control Rack...

Page 19: ...19 Notes www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30...

Page 20: ...Printed in the USA No VGMNL070319 1200CM www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30...

Reviews: