7-4
SurgiStat II Service Manual
Understanding Error Codes and Audio Tones
The SurgiStat II Generator includes automatic, continual self-diagnostics. If the
diagnostics detect an error, the system displays an error code, sounds an audible
tone, and deactivates the output power.
Any errors detected will shut down the RF output power.
Most error codes result from faults in accessories attached to the unit. The
following table lists the error codes, describes the error, and recommends actions
to take to resolve the error.
Notice
Internal firmware self-diagnostics continually monitor unit operation to ensure
proper and safe performance.
Error Code
Description
Recommended Action
F1
on the Cut/
Blend display
F1
on the
Desiccate/
Fulgurate/Bipolar
display
This is a start
activation type of
error and is the
result of powering
On while pressing
the pencil
activation button.
Handswitch or monopolar footswitch
cut pedal may be stuck
Handswitch or monopolar footswitch
coag pedal may be stuck
1.
Turn off, then turn on the generator. Do not press
buttons or accessory activation devices during the
self-test.
2.
If the alarm number reappears, disconnect all
accessories. Turn off, then turn on the generator
again.
3.
If the problem persists, replace the handpiece or
footswitch and repeat Step 1.
4.
If the alarm number reappears, record the number
and contact your Valleylab Service Center.
F2
Simultaneous
activation error
Cut and Coag buttons activated
simultaneously (pencil or footswitch)
The unit does not allow simultaneous activation of the
Cut and Coagulation modes. Release either the Cut or
Coag button on the handpiece, or the Cut or Coag pedal
on the footswitch.
F3
Foot activation
error
Monopolar handpiece activated while
in Bipolar mode
Release pressure on the handpiece buttons. Use the
footswitch to activate bipolar, or switch to monopolar
mode.
E4
High voltage
error
Line voltage error (Line voltage too
high; that is, the power supply is 30%
above the nominal value)
1.
Turn the unit off.
2.
Verify that the unit is connected to the correct line
voltage.
Summary of Contents for SurgiStat II
Page 1: ...Service Manual SurgiStat II Electrosurgical Generator ...
Page 24: ...2 12 SurgiStat II Service Manual ...
Page 46: ...3 22 SurgiStat II Service Manual Figure 3 20 Bipolar mode waveform ...
Page 60: ...5 6 SurgiStat II Service Manual ...
Page 64: ...6 4 SurgiStat II Service Manual ...
Page 74: ...7 10 SurgiStat II Service Manual ...