6-2
SurgiStat II Service Manual
Cleaning the Unit
Clean the generator after each use. Follow the procedures approved by your
institution or use a validated infection control procedure.
1.
Turn off the generator. Unplug the power cord from the wall outlet.
2.
Thoroughly wipe all surfaces of the generator and power cord with a mild
cleaning solution or disinfectant and a damp cloth.
Performing Periodic Inspection
Every six months, visually inspect the generator for signs of wear or damage. In
particular, look for any of the following problems:
•
Damage to the power cord
•
Damage to the power cable receptacle
•
Obvious damage to the unit
•
Damage to any receptacle
•
Accumulation of lint or debris in or around the unit
Warning
Electric Shock Hazard
Always turn off and unplug the generator before
cleaning.
Caution
Do not allow fluids to enter the generator chassis.
Do not sterilize the generator.
Notice
Do not clean the generator with abrasive cleaning or disinfectant compounds,
solvents, or other materials that could scratch the panels or damage the
generator.