8-4-4. 0.1 Amp Adjustment
Connect the 10 ohm standard resistor to the 4300B using four-wire connections. Select the 2V and .1A
ranges. Adjust RV104 for a reading of 10000m
Ω
.
8-4-5. 10mA Adjustment
Connect the 100 ohm standard resistor to the 4300B using four-wire connections. Select the 2V and 10mA
ranges. Adjust RV105 for a reading of 100.00
Ω
.
8-4-6. 1mA Adjustment
Connect the 1000 ohm standard resistor to the 4300B using four-wire connections. Select the 2V and 1mA
ranges. Adjust RV106 for a reading of 1000.0
Ω
.
8-4-7. 0.1mA Adjustment
Connect the 10000 ohm standard resistor to the 4300B using four-wire connections. Select the 2V and .1mA
ranges. Adjust RV107 for a reading of 10000
Ω
.