© 2019 Valeport Ltd
Data Requests and Output Formats
Page 15
·
The data separator is a tab (this may be altered if required).
·
Data is presented in the order: Pressure, Temperature, SV / Conductivity
·
For the miniSVP and miniCTD, pressure data format is dependent on sensor range, and may be any of the
following. Leading zeroes are included, so it is a fixed length string:
-
PPPP.P (e.g. 0123.4 dBar)
-
PPP.PP (e.g. 012.34 dBar)
-
PP.PPP (e.g. 12.345 dBar)
·
For the miniTIDE, pressure data is a fixed length string, 3 decimal places, including leading zeroes.
·
The temperature data (miniSVP & miniCTD) is given to 3 decimal places. Value is in °C and leading zeroes are
included; signed if negative:
-
21.456
-
02.769
-
-01.174
·
Sound Velocity (miniSVP) is given in m/s, as a fixed length string with 3 decimal places. In air, the sensor reads
0000.000
·
Conductivity (miniCTD) is given in mS/cm, as a fixed length string with 3 decimal places, and leading zeroes if
appropriate.