Hyperion Optical Sensors - Operating Manual:
MANUAL-1013637447-3 | issue: 1.6
Page | 5
©
2022
– Valeport Ltd
2.1.3 Phycocyanin
Performance
Excitation:
590 nm
Detection:
650 nm
Dynamic Range:
0-9 000 ppb
(Pre SN 75100: 2 gain settings, 0-45, 0-9000 (software controlled))
Instrument Detection limit:
<2 ppb*
Actual Detection limit:
2.08 ppb**
Linearity:
0.99 R
2
Response Time:
0.03 to 2 s
* 3x SD in RO water
** Calibrated against Phycocyanin in water\Phosphate buffer solution
2.1.4 Rhodamine WT
Performance
Excitation:
520 nm
Detection:
650 nm
Dynamic Range:
0-1000 ppb
(pre SN 75100: 2 gain settings, 0-50, 0-1000 (software controlled))
Instrument Detection limit:
<0.01 ppb*
Actual Detection limit:
0.06 ppb**
Linearity:
0.99 R
2
Response Time:
0.03 to 2 s
* 3x SD in RO water
** Calibrated against Rhodamine WT solution
2.1.5 Sulforhodamine B
Performance
Excitation:
520 nm
Detection:
650 nm
Dynamic Range:
0-1000 ppb
Instrument Detection limit:
<0.03 ppb*
Actual Detection limit:
<1 ppb**
Linearity:
0.99 R
2
Response Time:
0.03 to 2 s
* 3x SD in RO water
** Calibrated against Sulforhodamine B solution