10.4 Flow-Through Adapter Dimensions
shows the dimensions of the MGP261 flow-through adapter 258877 with
MGP261 installed into the adapter. The dimension figure also shows the measurement (from
the edge of the adapter mounting plate to the edge of the connection box) that can be used
to verify that the probe has been inserted to correct installation depth. The dimensions are
given in millimeters and [inches].
415[16.33]
130[5.1]
198[7.8]
G 1/8
Ø 55 [21.7]
Ø76 [29.9]
186 (±1 mm) [7.32]
Distance at correct installation depth
mm
[in]
Figure 29 Flow-Through Adapter Dimensions with MGP261 Probe
shows the dimensions and screw hole sizes of the flow-through adapter
mounting plate.
Ø 10
[0.39]
90 [3.54]
14.4 [0.56]
R20
Ø 6.2
[0.24]
9.1
[0.36]
39.1
[1.53]
12.5
[0.49]
30
[1.18]
70
[2.75]
mm
[in]
74
[2.91]
48.2
[1.9]
30
[1.18]
9.1
[0.36]
Figure 30 Flow-Through Adapter Mounting Plate Dimensions