3. Select
External
from either the
Pressure compensation mode
or the
Temperature
compensation mode
dropdown list, and then select
Save
.
4. Open the
Analog input 1
menu and configure the scaling of the external sensor input. See
Configuring Input from External Analog Pressure or Temperature Sensor (page 47)
7.3.4 Configuring Input from External Analog Pressure or
Temperature Sensor
When you set either the
Temperature compensation mode
or
Pressure compensation mode
to use measurement from an
External
sensor, you can configure the scale of the external
sensor input in the
Analog input 1
menu.
The
Input mode
and
Input parameter
selections are set automatically based on the external
compensation parameter selection.
Figure 22 Analog Input 1 Configuration Menu in Insight
1. In the
Measurement
menu, set either the pressure or temperature compensation mode as
External
to use input from a connected sensor. See
Using Compensation Received from
External Temperature or Pressure Sensor (page 46)
You can receive the compensation value from an external sensor for only one
parameter (temperature or pressure) at a time.
2. Store the setting with
Save
.
3. In the
Analog input 1
menu, enter the scaling for the temperature or pressure input from
the external sensor.
Chapter 7 – Environmental Compensation
47