3.4.1 Wiring Diagram
ZONE 1
HAZARDOUS AREA
ZONE 0
+
−
Vin +
Vin −
D +
D −
3 −
3 +
2 −
2 +
1 −
1 +
P/T
+
−
CONTROLLER
SAFE AREA
MGP261
EXTERNAL SENSOR
(PRESSURE OR TEMPERATURE)
GND
RS-485 +
RS-485 −
RSGND
4 ... 20 mA OUTPUT
S
4 ... 20 mA Ex ia INPUT
SHIELD
RS-485
Iout1
Iout2
Iout3
Iin
+ −
24 VDC
POWER SUPPLY FOR
ANALOG OUTPUTS
+
−
24 VDC
POWER SUPPLY
FOR PROBE
GROUNDING
TERMINAL
GROUNDING
RAIL
FOR POSSIBLE UNUSED WIRES;
INTERNALLY CONNECTED WITH
PROBE GROUNDING TERMINAL
EQUIPOTENTIAL
BONDING
Figure 13 MGP261 Wiring Diagram
Chapter 3 – Installation
29