3. When the measurement has stabilized, click the
Reference value, point 1
text box and
enter the known temperature of the calibration point. Press
ENTER
or click outside the
text box when done.
4. Check that the measured value for point 1 is automatically inserted.
5. Check the difference between the reference and the measured value. Very large
differences may be due to insufficient stabilization time or unsuitable calibration setup.
a. If you want to apply the adjustment you have made, select
Activate adjustment
.
b. To exit without taking the adjustment in use, select
Close
.
6. After completing the adjustment, update the information in the
Calibration Information
tab.
Chapter 8 – Calibration
57