User's Guide ______________________________________________________________________
116 _________________________________________________________________ M210667EN-C
0808-004
Figure 74
Fastening Screws of Measurement Unit
6.
Place the measurement unit on top of the inner mast tube (windows
pointing left). Do not use the air duct on top of the unit as a handle,
carry the measurement unit at the enclosure sides!
7.
The connection cable must be routed below the small junction box
near the blower.
8.
Make sure that the measurement unit has completely settled on the
inner mast tube of the transmitter mast. The air duct inlet on top of
the measurement unit must be placed in the same level than the
outlet of the blower. See Figure 75 below.
0810-077
Figure 75
Height of Measurement Unit
9.
Turn the measurement unit roughly into a parallel position with the
support unit (housing edges parallel to the outline of the support
unit). The measurement unit must have a distance of approximately
8 mm to all surrounding parts of the support unit. See Figure 76 on
page 117.