User's Guide ______________________________________________________________________
48 __________________________________________________________________ M010056EN-G
Lower Half of Display
The output quantities metric/non metric DIP switch can be used to
select between °C and °F.
It is possible to check the pressure setting of the transmitter by pressing
button
Dn
. To return to the temperature reading, press button
Dn
again.
Selecting Analog Outputs
You can select the output quantities for channels 1 and 2 by turning the
Calibration Enabled/Disabled
DIP switch of the transmitter to
position
Enabled
(up) and the three output selection DIP switches to
position special (all up).
Press button
C
on the display cover: text "SCAL" appears on the
display. Then press button
E
until the following display appears:
0505-288
The characters on the second line of the display in this menu correspond
to the quantities according to the table in section
.
Select the quantity for Ch 1 with buttons
Up
and
Dn
and acknowledge
the selection with button
E
.
If the transmitter is equipped with two analog channels, select the
quantity for Ch 2 in the same way, for example:
The characters used to indicate availability in the table above:
A
=
Available for HMT360 standard version
D
=
Available for HMT360 with optional calculations
F
=
Available for HMT360 Moisture and Temperature in oil
transmitter
H
=
Available for HMT360 Moisture and Temperature in Jet Fuel
transmitter