High End Adjustment (Two-Point Adjustment)
1.
After having made the low end adjustment, insert the probe into a
measurement hole of the NaCl (75 %RH) salt chamber in the
humidity calibrator. Please, note that the difference between the
two humidity references must be at least 30 %RH.
2.
The readings measured by the transmitter appear on the display, as
well as the text "Set Hi" on the lower left hand corner. If the
transmitter has no display, the LED indicator lights up.
3.
Wait at least 10 ... 15 minutes for the sensor to stabilize; use a
multimeter or the transmitter display to monitor the stabilization.
4.
Press buttons
Up
and
Dn
to adjust the display or multimeter
reading (when using a multimeter calculate the current value
corresponding to the reference humidity by using the equations
presented in
Calculating Correspondence of Current Values and
) and conclude the adjustment with
button
E
.
5.
Restore the DIP switch settings. Set the calibration DIP switch to
position disabled (down).
One-Point Temperature Adjustment
1.
Set the internal DIP switch
Calibration Enabled/Disabled
to
position
ON
and select T calibration by using switches 3 and 4 (see
section
DIP Switch Functions on page 45
2.
The temperature value measured by the transmitter appears on the
display as well as the text "Set Lo" on the lower left corner. If the
transmitter has no display, the LED indicator starts to blink.
NOTE
If you use serial commands, please refer to section
NOTE
Always use a high quality standard for adjusting the temperature.
NOTE