Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 49
0505-289
Press button
C
to exit the display command mode or continue by setting
the pressure.
Scaling Analog Outputs
Turn the internal
Calibration Enabled/Disabled
DIP switch of the
transmitter to upward position (on). Press button
C
on the display cover
and a text similar to the following appears:
0505-290
Numbers on the second line indicate the low end scaling of Ch 1
currently stored in the transmitter memory. The text "Set Lo" on the
lower left corner indicates that you can now change the low end scaling
with buttons
Up
and
Dn
. Acknowledge the value with button
E
and a
text similar to the following appears:
0505-291
Numbers on the second line indicate the high end scaling of Ch 1. The
text "Set Hi" on the lower left-hand corner indicates that you can now
change the high end scaling with buttons
Up
and
Dn
. Acknowledge
with button
E
.
NOTE