Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 47
0505-087
Adjust the pressure reading with buttons
Up
and
Dn
. Acknowledge the
value with button
E
. To complete the pressure setting, turn the dip
switch back to position disabled (down).
See the pressure conversion table on page
Selecting Output Quantities
Upper Half of Display
To modify the displayed quantity, turn the
Calibration Enabled/
Disabled
DIP switch of the transmitter to position
Enabled
(up). Select
the required quantity with button
Up
on the display cover, and
acknowledge the value with button
E
. To complete the selections, turn
the DIP switch back to position disabled (down).
Character
Quantity
Abbreviation
Availability
Metric Unit
Nonmetric Unit
0
relative humidity
RH
A, D
%RH
%RH
1
temperature
T
A, D, F, H
°C
°F
2
dewpoint
temperature
Td
D
°C
°F
3
absolute humidity a
D
g/m
3
gr/ft
3
4
mixing ratio
x
D
g/kg
gr/lb
5
wet bulb
temperature
Tw
D
°C
°F
6
mass
concentration of
water
ppmw
H
ppm
w
ppm
w
7
water activity
aw
H, F
0 ... 1
0 ... 1
8
relative humidity
of saturation
RS
H
%RS
%RS
9
saturation
temperature
Ts
H
°C
°F