Syntax: PUR x<cr>
where
x =
ON/OFF
Example:
>pur off
Chemical Purge : OFF
>pur on
Chemical Purge : ON
>
PUR Set Chemical Purge Interval
If the sensor is exposed to chemicals it is recommended to have the
chemical purge done at least once in 720 min (= 12 hours). In
applications where the chemical exposure is not likely, the interval
may be longer.
It is not recommended to change the other parameters than interval.
Syntax: PUR<cr>
Issue command PUR and type the new interval (in minutes) and press
Enter
five times to finish the command. The maximum interval is
14400 minutes (= 10 days).
Example:
>pur<cr>
Chemical Purge : ON
<cr>
Interval min : 1440 ?
<cr>
Duration s : 120 ?
<cr>
Settling s : 240 ?
<cr>
Power : 160 ?
<cr>
Temp diff : 0.50 ?
<cr>
>
NOTE
To activate the new interval setting immediately, reset the transmitter.
PURR Chemical Purge in Powerup
This command enables or disables the function, which turns on
chemical purge within 10 seconds from the powerup of the device.