USER'S GUIDE____________________________________________________________________
40 _______________________________________________________________ M210619EN-C
Format string consists of quantities and modifiers: use the quantity
abbreviations and modifiers presented in Table 5 below and in Table 6
below when selecting the output quantities.
Table 5
Quantity Abbreviations for FORM Command
Abbreviation Quantity
RH Relative
humidity
T Temperature
TDF Dewpoint/Frostpoint
TD Dewpoint
A Absolute
humidity
X Mixing
ratio
TW
Wet bulb temperature
PPM
Humid air volume/dry air volume
PW Water
vapor
pressure
PWS
Saturation water vapor pressure
H Enthalpy
NOTE
Only quantities selected when ordering the device can be selected for
an output quantity.
Table 6
Modifiers
Modifier Description
x.y
Length modifier (whole numbers and
decimal places)
#t Tabulator
#r Carriage
return
#n Line-feed
"" String-constant
U5
Unit field and length (unit can be
changed with command UNIT)
Examples:
>form "TD=" 5.2 TD #r#n<cr>
TD= -3.65
>form "TD=" TD U3 #t "TDF=" TDF U3 #r#n<cr>
TD= -4.0'C TDF= -3.6'C
>
Command FORM / returns the default output form:
>form /<cr>
>send<cr>
RH= 28.0 %RH T= 23.3 'C