Chapter 4 ________________________________________________________________ Operation
VAISALA____________________________________________________________________ 35
Calibration and adjustment (these are presented in
Calibration and Adjustment Commands
CRH
Relative humidity calibration
CT Temperature
calibration
LI
Revert factory calibration
FCRH
Relative humidity calibration after
sensor change
CTEXT
Set calibration information text
CDATE
Set calibration date
ACAL
Analog output calibration
Chemical purge
PUR [ON/
OFF
]
Enable or disable automatic
chemical purge
PURGE
Activate chemical purge
PURR [ON/
OFF
]
Enable or disable automatic
chemical purge in power up
Sensor heating
XHEAT [ON/OFF]
Enable or disable sensor heating
Measurement Output
R Start Continuous Outputting
Syntax: R<cr>
Starts output of measurements to the peripheral devices (RUN-mode);
the only command that can be used is S (stop).
The output mode can be changed with command FORM.
Example:
>r<cr>
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C