Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 51
Other Commands
Set Analog Output Mode
Use the
AMODE
command to show or set the operation mode of the
analog output. This command is not in use in the digital-only HMP110
probe option.
AMODE
[
ch1
] [
ch2
]<cr>
where
ch1
= Analog output mode for channel 1, range 0 ...3.
The options are:
0 (0 ... 1 V)
1 (0 ... 2.5 V)
2 (0 ... 5 V)
3 (1 ... 5 V)
ch2 = Analog output mode for channel 2, range 0 ...3.
The options are the same as for channel 1.
Example (show current output modes):
amode
Ch1 output : 0 ... 1 V
Ch2 output : 0 ... 1 V
Example (set channel 1 to 0 ... 1 V and channel 2 to 0 ... 5 V):
amode 0 2
Ch1 output : 0 ... 1 V
Ch2 output : 0 ... 5 V