User's Guide _______________________________________________________________________
42 ___________________________________________________________________ M211060EN-H
View Serial Number
Use the
SNUM
command to view the serial number of the probe.
SNUM
<cr>
Example:
snum
Serial number : H3640004
View Software Version
Use the
VERS
command to display the software version of the probe.
VERS
<cr>
Example:
vers
HMP63 / 1.0.4
Serial Line Output Commands
Start Measurement Output
Use the
R
command to start the continuous outputting of measurement
values as an ASCII text string to the serial line.
For HMP60 and HMP110, the output always includes readings for
temperature, RH and Td. For HMP110T, the output includes only
temperature.
R
<cr>
Example (HMP60 and HMP110):
r
T= 22.6 'C RH= 22.8 %RH Td= 0.3 'C
T= 22.6 'C RH= 22.5 %RH Td= 0.2 'C
T= 22.6 'C RH= 22.5 %RH Td= 0.2 'C
…
Example (HMP110T):
r
T= 22.6 'C
T= 22.6 'C
T= 22.6 'C
…