Clear Adjustment of T Measurement
Use the
CTCLR
command to clear the adjustment of temperature
measurement that has been done using the
CT
command.
CTCLR
<cr>
Example:
ctclr
OK
View User Adjustment Parameters
Use the
L
command to view the current user adjustment parameters. This
command is useful for checking the currently applied customer
calibration.
L
<cr>
The output values are as follows:
- Cp offset and gain: capacitance, calibrated using the
CRH
command
- T offset and gain: calibrated using the
CT
command
Example (shows default values, no user calibration done):
l
Cp offset : 0.0000
Cp gain : 1.0000
T offset : 0.0000
T gain : 1.0000