Table 26 HMP63 mechanical specifications
Property
Description/Value
IP rating
Cable connector
4-pin M8 (IEC 60947-5-2)
Materials
Body
PC/ABS blend
Grid filter
PC (glass reinforced)
Cable
Polyurethane or FEP
Weight
Probe
9 g (0.3 oz)
Probe with 0.3 m (1 ft) cable
20 g (0.7 oz)
1)
Not applicable with the plastic grid filter.
9.3 HMP110 specifications
Table 27 HMP110 measurement performance
Property
Description/Value
Relative humidity
Measurement range
0 … 100 %RH
Accuracy:
at 0 … +40 °C (+32 … +104 °F)
±1.5 %RH (0 … 90 %RH)
±2.5 %RH (90 … 100 %RH)
at −40 … 0 °C (−40 … +32 °F) and +40 … +80 °C
(+104 … +176 °F)
±3.0 %RH (0 … 90 %RH)
±4.0 %RH (90 … 100 %RH)
Factory calibration uncertainty
at +20 °C (+68 °F)
±1.1 %RH (0 … 90 %RH)
±1.8 %RH (90 … 100 %RH)
Humidity sensor types
HUMICAP
â
180R
HUMICAP
â
180V
Stability
±2 %RH over 2 years
Temperature
Measurement range
−40 … +80 °C (−40 … +176 °F)
Accuracy (probes with analog output):
Chapter 9 – Technical data