background image

7. Maintenance

7.1 Periodic maintenance

The humidity measurement accuracy of the HMP60 and HMP110 series probes should be
calibrated yearly. When calibration indicates that accuracy is not within specification:

• HMP60 and HMP63: change the INTERCAP

â

 sensor.

• HMP110 and HMP113: adjust the measurement yourself, or have it adjusted at Vaisala.

Light cleaning of the probe and replacement of the filter should be done only when necessary.

7.1.1 Cleaning

The probe body can be wiped clean with a soft, lint-free cloth moistened with mild detergent.
Do not use solvents or compressed air.

Note that wiping the membrane filter or stainless steel sintered filter may block its pores
and/or deposit residue on the filter. If the filter is heavily contaminated, replace it.

7.1.2 Replacing filter

The filter on the probe should be replaced when it is damaged or dirty.

1. Turn the filter counter-clockwise to loosen it.

2. Remove the filter from the probe. Be careful not to touch the sensors with the filter.

Without the filter in place, the sensors are easily damaged – handle the probe carefully.

3. Install a new filter on the probe, and tighten it so it is finger-tight. Make sure the filter sits

straight and meets the threads properly.

Order new filters from Vaisala.

More information

Filter options (page 13)

Spare parts and accessories (page 83)

Maintenance and calibration services (page 95)

7.2 Calibration procedure

To calibrate your HMP60 or HMP110 series probe, you need a known stable humidity or

temperature reference, and a way to read the output of the probe (Vaisala Insight software,

analog output, serial output, the MI70 indicator, or the HM40 meter). As a humidity reference

you can use, for example, the Vaisala Humidity Calibrator HMK15.

The probe can be connected to a PC with Vaisala USB cable 219690, and to the MI70 indicator

with connection cable 219980. The HM40 meter does not need a specific connection cable.

Chapter 7 – Maintenance

55

Summary of Contents for HMP110 series

Page 1: ...M211060EN J User Guide Vaisala Humidity and Temperature Probes HMP60 and HMP110 Series ...

Page 2: ...nal English versions In ambiguous cases the English versions are applicable not the translations The contents of this document are subject to change without prior notice Local rules and regulations may vary and they shall take precedence over the information contained in this document Vaisala makes no representations on this document s compliance with the local rules and regulations applicable at ...

Page 3: ...tory compliance 21 3 Installation 22 3 1 Mounting HMP60 HMP110 and HMP110T probes 22 3 1 1 Probe assembly with duct installation kit 22 3 1 2 Drilling instructions for duct installation kit 23 3 2 Mounting HMP63 and HMP113 probes 23 3 3 Wiring 24 3 3 1 Wiring multiple digital devices 25 3 3 2 Wiring with the loop power converter 26 3 3 3 Power supply requirements 27 4 Modbus communication 29 4 1 U...

Page 4: ...measurement 47 6 9 4 Clear adjustment of T measurement 48 6 9 5 View user adjustment parameters 48 6 10 Other commands 48 6 10 1 Set analog output mode 48 6 10 2 Set analog output parameters and scaling 49 6 10 3 Set analog output error indication level 50 6 10 4 Extend analog output range 51 6 10 5 Extend maximum RH reading 51 6 10 6 Display command list 52 6 10 7 Display the currently active err...

Page 5: ... 69 9 Technical data 70 9 1 HMP60 specifications 70 9 2 HMP63 specifications 72 9 3 HMP110 specifications 75 9 4 HMP113 specifications 78 9 5 Probe dimensions 81 9 5 1 HMP60 HMP110 and HMP110T dimensions 81 9 5 2 HMP63 and HMP113 dimensions 82 9 6 Spare parts and accessories 83 Appendix A Modbus reference 85 A 1 Communication settings 85 A 2 Function codes 85 A 3 Data encoding 86 A 3 1 32 bit floa...

Page 6: ...e 11 Assembly of the probe with duct installation kit 22 Figure 12 Drilling instructions 23 Figure 13 M8 4 pin male connector 24 Figure 14 Wiring multiple devices using local power supply 26 Figure 15 Wiring multiple devices using common power supply 26 Figure 16 Wiring with the loop power converter module 27 Figure 17 Connecting probe to Insight 31 Figure 18 PuTTY terminal application 36 Figure 1...

Page 7: ...Table 20 HMP60 inputs and outputs 71 Table 21 HMP60 mechanical specifications 72 Table 22 HMP63 measurement performance 72 Table 23 HMP63 calculated parameters 73 Table 24 HMP63 operating environment 74 Table 25 HMP63 inputs and outputs 74 Table 26 HMP63 mechanical specifications 75 Table 27 HMP110 measurement performance 75 Table 28 HMP110 calculated parameters 76 Table 29 HMP110 operating enviro...

Page 8: ...s status data registers read only 89 Table 44 Error codes on Modbus interface 90 Table 45 Modbus test registers read only 90 Table 46 Device identification objects 91 HMP60 and HMP110 Series User Guide M211060EN J 6 ...

Page 9: ...echnical data chapters Modbus reference page 85 Document format and layout update quality improvements and content reorganization M211060EN H August 2017 Previous version Removed the separate HMP110D digital output model digital output is now a selectable option when ordering the HMP110 model either analog or digital only HMP110 selected when ordering Added information on HUMICAP 180V sensor optio...

Page 10: ...60 and HMP110 Series Multilingual Quick Guide M210297EN Vaisala Handheld Humidity and Temperature Meter HM70 User Guide M211080EN Vaisala Mounting Flange for Humidity Probes Quick Reference Guide M211106EN Vaisala Loop Power Converter Quick Reference Guide 1 3 Documentation conventions Warning alerts you to a serious hazard If you do not read and follow instructions carefully at this point there i...

Page 11: ...â are registered trademarks of Vaisala Oyj Modbusâ is a registered trademark of Schneider Automation Inc Windowsâ is either a registered trademark or trademark of Microsoft Corporation in the United States and other countries All other product or company names that may be mentioned in this publication are trade names trademarks or registered trademarks of their respective owners Chapter 1 About th...

Page 12: ...2O2 for example vaporized hydrogen peroxide VHP is present HMP110 series probes can also be ordered with the HUMICAPâ 180V catalytic sensor HMP110 series probes require calibration after sensor replacement This can be done with Vaisala Insight PC software or on the serial line using the optional Vaisala USB cable Table 3 Parameters measured by HMP60 and HMP110 series Parameter Abbreviation Metric ...

Page 13: ...ormance as HMP110 One analog output channel 4 HMP110REF IP65 rated stainless steel probe Does not measure instead outputs constant humidity and temperature readings on serial line Useful for validating installations of HMT120 and HMT130 transmitters for example Digital output only 5 HMP63 IP54 rated probe with PC ABS plastic housing and INTERCAPâ sensor Lightweight probe with faster thermal respon...

Page 14: ...wer converter 4 20 mA separate module compatible with humidity accuracy only Shielded 0 3 m and 3 0 m 1 ft and 9 8 ft connection cables with threaded connector for probe connection open end wires on the other end Plastic M12 installation nuts for HMP60 HMP110 and HMP110T Plastic locking bushing for HMP63 and HMP113 for use with Vaisala products for example the HM40 handheld meter 2 2 1 Output opti...

Page 15: ... 2 Yes Modbus RTU Yes Yes Yes Yes Yes Yes Yes 1 4 20 mA with separate loop power converter compatible with humidity accuracy only 2 Temporary service access with serial line only 2 2 2 Filter options 1 2 3 4 5 Figure 2 Filters for HMP60 HMP110 and HMP110T 1 Plastic grid filter fastest response time 2 Membrane filter pore size 0 2 µm 3 Stainless steel sintered filter pore size 38 µm 4 PTFE membrane...

Page 16: ... 3 Installation accessories optional For item codes see Spare parts and accessories page 83 2 3 1 Probe mounting clamp The optional mounting clamp makes it easy to install the probe on the wall of the measurement environment The probe can be detached for calibration simply by loosening the lower screw Figure 4 Probe mounting clamp in use The probe mounting clamp is delivered in 2 parts that must b...

Page 17: ...1 Align the slots on the clamp parts 2 Slide the lower clamp part over to the bottom end of the upper part Chapter 2 Product overview 15 ...

Page 18: ... that you can use to hold the probe in a through wall installation The flange is a general purpose mounting accessory for Ø 12 mm probes and comes with a sealing plug for coaxial cables that is not needed when you use the flange with HMP60 and HMP110 series probes For more information see Mounting Flange for Humidity Probes Quick Reference Guide M211080EN Figure 5 Probe mounting flange HMP60 and H...

Page 19: ...113 to compatible Vaisala instruments using a plastic locking bushing that is placed over the probe The bushing has a M15 1 thread It is compatible with HMT120 and HMT130 transmitters and HM40 handheld meter Figure 6 HMP113 with plastic locking bushing Chapter 2 Product overview 17 ...

Page 20: ... assemble the probe to the duct installation kit slide the probe head through the hole and attach the flange to the duct wall with 4 screws Ø 24 0 94 1 2 mm in Figure 7 Probe installation with the duct installation kit 1 Tension screw 2 Installation depth Adjust the depth and lock in place with the tension screw More information Mounting HMP60 HMP110 and HMP110T probes page 22 HMP60 and HMP110 Ser...

Page 21: ...nalog output mode The desired parameter is on channel 1 which must be scaled to 0 2 5 V Figure 8 Loop power converter More information Wiring with the loop power converter page 26 2 3 6 Cables Connection cables have a straight threaded female M8 connector on one end and open wires on the other end You can use other compatible M8 series cables Figure 9 Cable with threaded connector Chapter 2 Produc...

Page 22: ...the unit Improper modification can damage the product or lead to malfunction CAUTION Before you connect an HMP60 or HMP110 series probe to a device it is recommended to power off the device 2 4 1 ESD protection Electrostatic Discharge ESD can damage electronic circuits Vaisala products are adequately protected against ESD for their intended use However it is possible to damage the product by deliv...

Page 23: ... electromagnetic compatibility of HMP60 HMP110 HMP110T and HMP110REF has been tested according to the following product family standards EN 61326 1 Electrical equipment for measurement control and laboratory use EMC requirements for use in industrial locations EN 55022 Class B Information technology equipment Radio disturbance characteristics Limits and methods of measurement The electromagnetic c...

Page 24: ...ng clamp to hold the probe on a wall Use the probe mounting flange to hold the probe in a through wall installation Avoid placing the probe in a place where condensation can run onto the sensor 3 1 1 Probe assembly with duct installation kit 1 2 3 Figure 11 Assembly of the probe with duct installation kit 1 HMP60 HMP110 or HMP110T probe 2 Duct installation kit 3 Probe cable 1 Slide the probe cable...

Page 25: ...m drill bit to drill the holes for the mounting screws four ST4 2 16 C Z DIN 7981 screws 3 2 Mounting HMP63 and HMP113 probes HMP63 and HMP113 probes do not have a thread on the probe body For a convenient installation use the optional installation accessories Use the probe mounting clamp to hold the probe on a wall Use the probe mounting flange to hold the probe in a through wall installation If ...

Page 26: ... 2 5 V 8 28 VDC Vout 0 5 1 5 V 2 Channel 1 RH Td T 0 1 2 5 5 V 1 5 V 3 GND AGND 4 Channel 2 RH Td T 0 1 2 5 5 V 1 5 V HMP110T has no output on channel 2 Pin out of HMP110REF HMP110 digital 1 5 28 VDC 2 RS 485 B 3 GND 4 RS 485 A The grounding method depends on the probe and the installation type Table 5 Grounding methods Probe Grounding method HMP63 HMP113 Vaisala recommends to use a shielded cable...

Page 27: ...and the analog output stabilizes The delay depends on the output type and on the operating voltage that is supplied to the probe Probes with analog output 4 s at operating voltage 13 5 16 5 VDC 2 s at other valid operating voltages Probes with digital output 2 s 3 3 1 Wiring multiple digital devices The maximum number of HMP60 HMP110 probes that you can connect to a system over the RS 485 interfac...

Page 28: ...D Figure 14 Wiring multiple devices using local power supply 1 VDC 3 GND 2 B 4 A HMP60 HMP110 M8 CONNECTOR 5 28 VDC COM 5 28 VDC COM A B A B TO PREVIOUS DEVICE TO NEXT DEVICE Figure 15 Wiring multiple devices using common power supply 3 3 2 Wiring with the loop power converter To use the loop power converter module with an HMP60 or HMP110 series probe the probe must be in the analog output mode Th...

Page 29: ...e loop power converter module For more information see the Loop Power Converter Quick Reference Guide M211106EN More information Output options page 12 3 3 3 Power supply requirements The operating voltage for the HMP60 and HMP110 series probes must be in the following range Table 6 Operating voltage ranges HMP60 HMP63 HMP110 analog HMP113 HMP110T HMP110REF HMP110 digital 5 28 VDC Vout 0 1 0 2 5 V...

Page 30: ...impedance loads on the signal outputs increase the current consumption by up to 0 5 mA High impedance loads are recommended to minimize warming of the probe Frequent interrogation of the probe using the RS 485 interface will also increase current consumption from the average value More frequent interrogation than once per second is not recommended HMP60 and HMP110 Series User Guide M211060EN J 28 ...

Page 31: ...de Modbus RTU Use Vaisala Insight software to change the Modbus serial communication settings if needed Download Vaisala Insight software at www vaisala com insight The instrument must be switched off and on before the communication setting changes take effect More information Accessing serial line command interface from analog or Modbus mode page 36 Modbus reference page 85 4 1 Using multiple dev...

Page 32: ...P60 HMP110 probes that can be connected If other devices require the use of termination HMP60 HMP110 series probes must be connected using an RS 485 repeater More information Wiring multiple digital devices page 25 HMP60 and HMP110 Series User Guide M211060EN J 30 ...

Page 33: ...oftware at www vaisala com insight The probe can be connected to Vaisala Insight software using a Vaisala USB cable item code 219690 HMP60 and HMP110 probes support Insight from probe software version 2 1 4 onwards 5 1 Connecting to Insight software Computer with Microsoft Windowsâ operating system and Vaisala Insight software installed USB connection cable item code 219690 When connecting several...

Page 34: ...the Insight software 2 Connect the USB cable to a free USB port on the PC 3 Connect the probe to the USB cable 4 Wait for Insight software to detect the probe HMP60 and HMP110 Series User Guide M211060EN J 32 ...

Page 35: ...ore you can use the cable you must install the provided USB driver on your PC The Vaisala USB cable is not designed for permanent installations When using the USB cable no separate power unit is needed The probe is powered through the USB port For permanent interfacing to a host system use a shielded cable with a threaded connector The probe does not echo typed characters back to the terminal scre...

Page 36: ...download the latest driver from www vaisala com software 3 Execute the USB driver installation program setup exe and accept the installation defaults The installation of the driver may take several minutes 4 After the driver has been installed connect the USB cable to a USB port on your PC Windows detects the new device and uses the driver automatically 5 The installation has reserved a COM port f...

Page 37: ...hich port the USB cable is using select Start Vaisala Vaisala USB Instrument Finder 4 Check that the other serial settings are correct for your connection and change if necessary 5 To open the connection window and start using the serial line select Open If PuTTY is unable to open the serial port you selected it shows an error message instead If this happens restart PuTTY and check the settings 6 ...

Page 38: ...t 2 Open the terminal program and open a connection to the corresponding COM port using the default settings 19200 8 N 1 no flow control 3 Select the Serial settings category and check that the Serial line to connect to field contains the correct COM port To check which port the USB cable is using select Start Vaisala Vaisala USB Instrument Finder 4 To open the connection window and start using th...

Page 39: ...e or lowercase In the command examples the keyboard input by the user is in bold type The notation cr refers to pressing the carriage return ENTER key on your computer keyboard Press ESC to clear the command buffer before starting to enter commands Table 10 List of serial commands software version 2 4 0 Command Description Output information about the device AERR Set analog output error level AMOD...

Page 40: ...UNIT Select metric or non metric output units VERS View software version of the probe Table 11 Additional commands for probes with RS 485 output Command Description Output information about the device in POLL mode ADDR 0 255 Set the probe address for POLL mode CLOSE Close the temporary connection Back to POLL mode OPEN 0 255 Open a temporary connection to a POLL mode device SERI baud p d s User Po...

Page 41: ... 1 V Ch2 output 0 1 V Ch1 RH lo 0 00 RH Ch1 RH hi 100 00 RH Ch2 T lo 40 00 C Ch2 T hi 60 00 C If the probe is in POLL mode but a connection has not been opened using the OPEN command issue the command cr More information Set serial line settings page 42 6 6 2 View calibration information Use the CDATE command to view the calibration date and CTEXT to view the calibration info text Date format for ...

Page 42: ...Room 1 6 6 4 View order code Use the CODE command to view the order code that has been stored in the probe This command is useful if you need to order a new probe with the same options CODE cr Example code Order code A12A0A2B0 6 6 5 View serial number Use the SNUM command to view the serial number of the probe SNUM cr Example snum Serial number H3640004 6 6 6 View software version Use the VERS com...

Page 43: ...P110T r T 22 6 C T 22 6 C T 22 6 C Outputting the results continues in intervals issued with the command INTV You can stop the output by entering the S command 6 7 2 Stop measurement output Use the S command or press the ESC key to stop the continuous measurement output S cr 6 7 3 Output the measurement message once Use the SEND command to output the measurement values once If the probe is in POLL...

Page 44: ...e when the probe is reset or powered up SERI b p d s Syntax Description b baud rate 300 600 1200 2400 4800 9600 19200 38400 57600 p parity n none e even o odd d data bits 7 or 8 s stop bits 1 or 2 Example shows default settings seri Baud P D S 19200 N 8 1 6 8 2 Set serial interface mode Use the SMODE command to set the operation mode of the serial interface The new mode is applied when probe is re...

Page 45: ...VDIGI Special serial interface mode that is only used for interoperability with Vaisala devices such as HMT120 HMT130 and HM40 This mode is set at Vaisala for probes that are ordered for such use ANALOG No serial line analog outputs active For instructions on how to enter the serial line when in analog mode see Accessing serial line command interface from analog or Modbus mode page 36 Note that an...

Page 46: ...8 4 Set measurement filtering Use the FILT command to view or set the speed at which the latest measurement result is integrated into the humidity and temperature readings The command affects both analog output and serial line output FILT a aaa cr Syntax Description a aaa Range 0 001 1 0 1 0 No filtering the latest measurement is output without averaging 0 5 Average of the last 2 measurements 0 1 ...

Page 47: ...l line settings page 42 6 8 6 Set serial interface delay Use the SDELAY command to view or set the serial interface answer minimum delay SDELAY delay cr Syntax Description delay Range 0 255 Value corresponds to 4 milliseconds for example 5 0 020 second minimum answer delay Example sdelay Serial delay 30 sdelay 50 Serial delay 50 6 8 7 Set measurement units Use the UNIT command to view or set the m...

Page 48: ...ation with a single 50 RH reference will adjust the gain of the capacitance measurement 2 point calibration will adjust both offset and gain The first point requires a 50 RH humidity reference the second point must be 50 RH There must also be at least 30 percentage point difference between the references CRH reference cr This command is not available on the HMP110T When performing a 1 point calibr...

Page 49: ...alibration adjusts the offset for the measurement 2 point calibration adjusts offset and gain CT reference cr When performing a 1 point calibration you need to place the probe in a single temperature reference and wait for 20 40 minutes for the temperature to stabilize To apply the adjustment enter the CT command with the reference temperature as a parameter Example 1 point calibration ct 23 5 OK ...

Page 50: ...ly applied customer calibration L cr The output values are as follows Cp offset and gain capacitance calibrated using the CRH command T offset and gain calibrated using the CT command Example shows default values no user calibration done l Cp offset 0 00000000E 00 Cp gain 1 00000000E 00 T offset 0 00000000E 00 T gain 1 00000000E 00 6 10 Other commands 6 10 1 Set analog output mode Use the AMODE co...

Page 51: ...scaling Use the ASEL command to show or set the output parameters and scaling of the analog outputs This command is not in use in the digital only HMP110 probe option ASEL ch1 ch2 ch1low ch1high ch2low ch2high cr Syntax Description ch1 Output parameter for channel 1 The options are RH Relative humidity T Temperature Td Dew point temperature Tw Wet bulb temperature a Absolute humidity x Mixing rati...

Page 52: ... rh Ch1 T lo 20 00 C 40 Ch1 T hi 80 00 C 60 Ch2 RH lo 0 00 RH Ch2 RH hi 100 00 RH 6 10 3 Set analog output error indication level If the device is malfunctioning the analog output is set to a specified level This overrides the normal measurement output of the channel The default error level is 0 V or another value predefined by the customer when ordering the device You can set the level using the ...

Page 53: ...he parameter remains as before the extra range is used for additional measurement range in the wet end This command is not in use in the digital only HMP110 probe option AOVER ON OFF cr The following example illustrates how the analog output is affected Channel 1 outputs Td with voltage output 0 5 V 40 C 60 C After giving the AOVER ON command the range is 0 5 5 V 40 C 70 C Note that the 60 C T d p...

Page 54: ... V 0 100 to 0 1 V 0 120 rhlimit 120 Max RH 120 0 RH asel Ch1 T lo 20 00 C Ch1 T hi 80 00 C Ch2 RH lo 0 00 RH Ch2 RH hi 100 00 RH 120 More information Set analog output parameters and scaling page 49 6 10 6 Display command list Use the HELP command to display a list of the currently available commands HELP cr 6 10 7 Display the currently active errors Use the ERRS command to display the currently a...

Page 55: ...settings corrupted 51 SCOEFS check sum error Sensor coefficients corrupted 52 CURRENT check sum error Main configuration settings corrupted 53 Voltage error Supply voltage out of range 55 General flash failure w r Non volatile memory read write failure 57 Calibration certificate check sum failure Calibration certificate checksum mismatch 58 Refer to Error messages in Insight software page 67 for m...

Page 56: ...re version at reset reset HMP60 1 00 0 Example probe set to serial mode RUN will start to output measurement messages at reset reset T 23 6 C RH 20 2 RH Td 0 5 C T 23 6 C RH 20 2 RH Td 0 5 C T 23 3 C RH 20 2 RH Td 0 8 C 6 10 11 Restore factory settings Use the FRESTORE command to restore the factory settings to the probe All user settings including the user performed calibration corrections will b...

Page 57: ...e filter counter clockwise to loosen it 2 Remove the filter from the probe Be careful not to touch the sensors with the filter Without the filter in place the sensors are easily damaged handle the probe carefully 3 Install a new filter on the probe and tighten it so it is finger tight Make sure the filter sits straight and meets the threads properly Order new filters from Vaisala More information ...

Page 58: ...g to stabilize Do not touch the probe or breathe in its direction during this time If you are using the MI70 indicator or the HM40 meter you can follow the stabilization using the graph functionality If you are using the USB cable do not keep the cable connected to the PC during the stabilization period for best calibration accuracy 4 After the stabilization period With MI70 HM40 Switch MI70 HM40 ...

Page 59: ...r of the humidity calibrator 11 RH 3 Start a terminal application and set the correct connection settings The default serial settings are 19200 8 N 1 Remember to check which COM port the USB cable is using 4 Wait for 20 40 minutes for the humidity to stabilize 5 Start the terminal session and connect the USB cable to the probe If your probe is in the analog output mode press ENTER a few times to s...

Page 60: ...rence the second point must be 50 RH There must also be at least 30 RH difference between the references The following procedure uses the HMK15 Humidity Calibrator LiCl salt 11 RH is used as the first reference point NaCl 75 RH as the second 1 Connect the USB cable to the PC but do not connect it to the probe yet 2 Remove the filter from the probe and insert the probe in the LiCl salt chamber of t...

Page 61: ... Use the SEND command to verify the currently measured RH value send T 22 9 C RH 11 1 RH Td 8 0 C 10 Give the CRH command with the RH value of the humidity reference as a parameter crh 11 OK 11 After entering the correction unplug the USB cable from the probe Insert the probe in the NaCl salt chamber 75 RH and wait for 20 40 minutes for humidity and temperature to stabilize 12 Connect the USB cabl...

Page 62: ...e 19200 8 N 1 Remember to check which COM port the USB cable is using 4 Wait for 20 40 minutes for the temperature to stabilize 5 Start the terminal session and connect the USB cable to the probe If your probe is in the analog output mode you need to press ENTER a few times to start it in RS 485 mode 6 Verify that the connection works by giving the command If the probe does not respond with device...

Page 63: ...70 meter see the HM70 Handheld Humidity and Temperature Meter User Guide M210297EN 7 4 1 1 point adjustment of RH measurement using MI70 indicator HMP110 and HMP113 If you want to perform a 2 point adjustment instead of a 1 point adjustment use Vaisala Insight software or the serial line See Vaisala Insight software page 31 and 2 point adjustment of RH measurement HMP110 and HMP113 page 58 To perf...

Page 64: ...eturn to the basic display More information 2 point adjustment of RH measurement HMP110 and HMP113 page 58 7 4 2 1 point adjustment of temperature measurement using MI70 indicator HMP110 and HMP113 Temperature adjustment can be done if there is reason to believe that the adjustment is changed In a 1 point adjustment make sure the reference condition represents the measuring environment To perform ...

Page 65: ...ent mode and EXIT to return to the basic display 7 5 Repair maintenance 7 5 1 Changing the INTERCAP sensor HMP60 and HMP63 This procedure restores the humidity measurement accuracy of the probe No adjustment after the sensor change is needed To perform this procedure you need a new INTERCAPâ sensor It is also recommended that you replace the filter with a new one 1 Remove the filter from the probe...

Page 66: ...s likely that the accuracy can be restored by performing the adjustment procedure Follow this procedure to replace the humidity sensor of the HMP110 and HMP113 in case it has been damaged or normal adjustment is not sufficient to restore the measurement accuracy Calibration and adjustment of the humidity measurement is required after the sensor change To perform this procedure you need a new HUMIC...

Page 67: ...ure sensor 3 Pull out the old HUMICAPâ sensor and insert a new one Handle the new sensor by the plastic frame Do not touch the sensor plate CAUTION 4 Perform a 2 point adjustment of the RH measurement 5 Attach a new filter on the probe More information 2 point adjustment of RH measurement HMP110 and HMP113 page 58 Chapter 7 Maintenance 65 ...

Page 68: ...le to remove the errors contact Vaisala For contact information see Technical support page 95 Table 14 Troubleshooting table Problem or message Likely causes and solutions Measurement not working Any of the following errors active T meas error RH meas error Sensor s damaged or missing Open the filter and check Check the supply voltage of the probe Check the output mode of the probe using serial li...

Page 69: ...measurement outside the permissible value range Check for condensation on the probe and sensor Allow the probe and sensor to dry If the error is still active there may be a problem with the electronics Contact Vaisala More information Technical support page 95 8 3 Error messages in Insight software Table 15 Error messages in Insight software Error Likely cause Suggested solution Temperature measur...

Page 70: ... Non volatile memory read write failure 57 Calibration certificate checksum mismatch 58 In case of constant error contact Vaisala More information Technical support page 95 8 4 Error codes in MI70 Handheld Meter If several errors are active at the same time the MI70 shows the sum of currently active error codes The label hex might not be displayed in MI70 Refer to Error messages in Insight softwar...

Page 71: ... 200hex Main configuration settings corrupted 53 800hex Supply voltage out of range 55 2000hex Non volatile memory read write failure 57 4000hex Calibration certificate checksum mismatch 58 8 5 Unknown serial settings You can check the current serial settings of your probe by connecting to the serial line or with Vaisala Insight software More information Vaisala Insight software page 31 Accessing ...

Page 72: ...ange 40 60 C 40 140 F Accuracy at 10 30 C 50 86 F at 40 10 and 30 60 C 40 50 and 86 140 F 0 5 C 32 9 F 0 6 C 33 08 F Dew point temperature Measurement range 40 60 C 40 140 F Typical accuracy at 32 40 C 0 140 F when dew point depression 1 15 C 59 F 2 C 35 6 F at 40 0 C and 40 60 C 40 32 F and 104 140 F when dew point depression 10 C 50 F 3 C 37 4 F Analog outputs Accuracy at 20 C 68 F 0 2 of FS Tem...

Page 73: ... 20 HMP60 inputs and outputs Property Description Value Power consumption 1 mA average max peak 5 mA Operating voltage 1 With 1 V 2 5 V output 5 28 VDC With 5 V output 8 28 VDC With loop power converter 8 28 VDC With digital output 5 28 VDC Start up time Probes with analog output 4 s at operating voltage 13 5 16 5 VDC 2 s at other valid operating voltages Probes with digital output 1 s Outputs 2 c...

Page 74: ...Body thread MI2x1 10 mm 0 4 in Cable connector 4 pin M8 IEC 60947 5 2 Materials Body Stainless steel AISI 316 Grid filter Chrome coated ABS plastic Cable Polyurethane or FEP Weight Probe 17 g 0 6 oz Probe with 0 3 m 1 ft cable 28 g 1 oz 1 Applicable with stainless steel sintered filter and PTFE sintered filter only 9 2 HMP63 specifications Table 22 HMP63 measurement performance Property Descriptio...

Page 75: ... 140 F when dew point depression1 15 C 2 C 35 6 F at 40 0 C and 40 60 C 40 32 F and 104 140 F when dew point depression 10 C 1 3 C 37 4 F Analog outputs Accuracy at 20 C 68 F 0 2 of FS Temperature dependence 0 01 of FS C 0 006 of FS F 1 Dew point depression ambient temperature dew point Table 23 HMP63 calculated parameters Measurement ranges Dew point temperature 40 60 C 40 140 F Wet bulb temperat...

Page 76: ...log output 4 s at operating voltage 13 5 16 5 VDC 2 s at other valid operating voltages Probes with digital output 1 s Outputs 2 channels 0 1 VDC 0 2 5 VDC 0 5 VDC 1 5 VDC 1 channel loop power converter separate module compatible with humidity accuracy only 4 20 mA Digital output optional RS 485 2 wire half duplex supports Modbus RTU External loads 0 1 V RL min 10 kΩ 0 2 5 V 0 5 V RL min 50 kΩ Out...

Page 77: ...10 specifications Table 27 HMP110 measurement performance Property Description Value Relative humidity Measurement range 0 100 RH Accuracy 1 2 at 0 40 C 32 104 F 1 5 RH 0 90 RH 2 5 RH 90 100 RH at 40 0 C 40 32 F and 40 80 C 104 176 F 3 0 RH 0 90 RH 4 0 RH 90 100 RH Factory calibration uncertainty at 20 C 68 F 1 1 RH 0 90 RH 1 8 RH 90 100 RH Humidity sensor types HUMICAPâ 180R HUMICAPâ 180V Stabili...

Page 78: ...point depression 15 25 C 59 77 F 1 C 33 8 F 2 C 35 6 F at 40 0 C 40 32 F and 40 80 C 104 176 F when when dew point depression 15 C 59 F 2 C 35 6 F Analog outputs Accuracy at 20 C 68 F 0 2 of FS Temperature dependence 0 01 of FS C 0 006 of FS F 1 Including non linearity hysteresis and repeatability 2 With HUMICAPâ 180V sensor accuracy is not specified below 20 C 4 F operating temperature 3 Dew poin...

Page 79: ... 28 VDC With 5 V output 8 28 VDC With loop power converter 8 28 VDC With digital output 5 28 VDC Start up time HMP110 probes with analog output 4 s at operating voltage 13 5 16 5 VDC 2 s at other valid operating voltages HMP110 probes with digital output 1 s Outputs 2 channels 0 1 VDC 0 2 5 VDC 0 5 VDC 1 5 VDC 1 channel loop power converter separate module compatible with humidity accuracy only 4 ...

Page 80: ...ed ABS plastic Cable Polyurethane or FEP Weight Probe 17 g 0 6 oz Probe with 0 3 m 1 ft cable 28 g 1 oz 1 Applicable with stainless steel sintered filter and PTFE sintered filter only 9 4 HMP113 specifications Table 32 HMP113 measurement performance Property Description Value Relative humidity Measurement range 0 100 RH Accuracy incl non linearity hysteresis and repeatability at 0 40 C 32 104 F 1 ...

Page 81: ...depression 15 C 59 F 1 when dew point depression 15 25 C 59 77 F 1 C 33 8 F 2 C 35 6 F at 40 0 C 40 32 F and 40 60 C 104 140 F when dew point depression 15 C 59 F 2 C 35 6 F Analog outputs Accuracy at 20 C 68 F 0 2 of FS Temperature dependence 0 01 of FS C 0 006 of FS F 1 Dew point depression ambient temperature dew point Table 33 HMP113 calculated parameters Property Description Value Measurement...

Page 82: ...alog output 4 s at operating voltage 13 5 16 5 VDC 2 s at other valid operating voltages Probes with digital output 1 s Outputs 2 channels 0 1 VDC 0 2 5 VDC 0 5 VDC 1 5 VDC 1 channel loop power converter separate module compatible with humidity accuracy only 4 20 mA Digital output optional RS 485 2 wire half duplex supports Modbus RTU External loads 0 1 V RL min 10 kΩ 0 2 5 V 0 5 V RL min 50 kΩ Ou...

Page 83: ...ABS blend Grid filter PC glass reinforced Cable Polyurethane or FEP Weight Probe 9 g 0 3 oz Probe with 0 3 m 1 ft cable 20 0 7 oz 1 Not applicable with the plastic grid filter 9 5 Probe dimensions 9 5 1 HMP60 HMP110 and HMP110T dimensions Ø 12 0 47 71 2 8 M12x1 Figure 19 HMP60 HMP110 and HMP110T dimensions in mm in Chapter 9 Technical data 81 ...

Page 84: ...5 2 HMP63 and HMP113 dimensions 79 5 3 13 24 5 0 96 3 5 0 14 Ø 14 0 55 Ø 12 0 47 Figure 21 HMP63 and HMP113 dimensions in mm in 68 2 68 M15x1 Ø 21 6 0 85 Figure 22 HMP63 and HMP113 with plastic locking bushing HMP60 and HMP110 Series User Guide M211060EN J 82 ...

Page 85: ...180V HUMICAP180V Sensor protection For HMP60 and HMP110 Plastic grid filter DRW010522SP Membrane filter DRW010525SP Stainless steel sintered filter HM46670SP PTFE membrane filter with stainless steel grid ASM212652SP PTFE sintered filter DRW244938SP For HMP63 and HMP113 Plastic grid filter DRW240185SP Plastic grid with membrane filter ASM210856SP Stainless steel sintered filter HM47280SP Porous PT...

Page 86: ...r converter 225979 USB cable for PC connection 219690 Connection cable for MI70 indicator 219980SP Connection cables with open wires 60 C 0 3 m 140 F 1 ft HMP50Z032SP 60 C 1 2 m 140 F 4 ft HMP50Z120 60 C 3 m 140 F 9 8 ft HMP50Z300SP 80 C 1 5 m 176 F 5 ft 225777SP 80 C 3 m 176 F 10 ft 225229SP 180 C 1 5 m 356 F 5 ft FEP 238025 180 C 3 m 356 F 10 ft FEP 226902SP 1 Connection cable 219980SP is also n...

Page 87: ... 0 Communication mode Modbus RTU You can use up to 32 probes on the same RS 485 line when the communication speed is 19200 bps or lower You must configure each probe on the line to have a different Modbus address RS 485 termination must not be used with HMP60 HMP110 series probes After power up it may take up to 2 seconds before the probe responds to any Modbus request A 2 Function codes Table 39 ...

Page 88: ...elease 1 0 This is also known as little endian or Modicon word order Despite the specification some Modbus masters may expect a big endian word order most significant word first In such case you must select word swapped floating point format in your Modbus master for the Modbus registers of the device A complete 32 bit floating point or 32 bit integer value should be read and written in a single M...

Page 89: ...tart from zero Register number 1 corresponds to address 0hex in the actual Modbus message Reading the wrong register s may result in correct looking values Check the reference documentation of your Modbus host PLC to verify which notation it uses for Modbus register addresses CAUTION A 4 1 Measurement data registers Table 41 Modbus measurement data registers read only Register number Address Regis...

Page 90: ...ng ratio 16 bit integer g kg 10 266 0109hex Wet bulb temperature 16 bit integer C 10 270 010Dhex Enthalpy 16 bit integer kJ kg 10 1 Only temperature output is available in probe model HMP110T A 4 2 Configuration registers Table 42 Modbus configuration data registers writable Register number Register address Register description Data format Unit Valid range Filtering 785 0310hex Measurement filteri...

Page 91: ...rted it might no longer be possible to communicate with the probe using Modbus You can override invalid communication settings in Vaisala Insight software Settings menu See Vaisala Insight software page 31 A 4 3 Status registers Table 43 Modbus status data registers read only Register number Address Register description Data format Note 513 0200hex Error status 16 bit integer 0000hex One or more e...

Page 92: ... read write failure 57 16384 Calibration certificate checksum mismatch 58 If several errors are active at the same time the Modbus error code register contains the sum of currently active error codes Refer to Error messages in Insight software page 67 for more information about the error states A 4 4 Test value registers Read the known test values from the test registers to verify the functionalit...

Page 93: ...0 Software version of the device 3 03hex VendorUrl http www vaisala com 4 04hex ProductName Vaisala HUMICAP R Humidity and Temperature Probe HMP113 5 05hex ModelName R00A0C1A0 Configuration code of the device Length and content of the code are model specific 128 80hex SerialNumber 1 J1140501 129 81hex CalibrationDate 1 2020 06 01 Calibration date in YYYY MM DD format Empty string if not set valid ...

Page 94: ...ster address 04hex Number of data bytes 00hex 7Ahex Value of first register least significant word 00hex Number of 16 bit registers to read 2 E1hex 02hex 41hex Value of second register most significant word D1hex Modbus RTU checksum F4hex 2Ahex 62hex Modbus RTU checksum silence for 3 5 bytes End of Modbus RTU frame 05hex silence for 3 5 bytes End of Modbus RTU frame Communication description Regis...

Page 95: ...ilence for 3 5 bytes End of Modbus RTU frame 3Ehex Value for the second register most significant word The response to a write function informs that the function was correctly received by the device It does not guarantee that the written value was accepted by the device for example in case of out of range values To verify that the value was really accepted by the device read the register value aft...

Page 96: ...HMP60 and HMP110 Series User Guide M211060EN J 94 ...

Page 97: ...anty Please observe that any such warranty may not be valid in case of damage due to normal wear and tear exceptional operating conditions negligent handling or installation or unauthorized modifications Please see the applicable supply contract or Conditions of Sale for details of the warranty for each product Technical support Contact Vaisala technical support at helpdesk vaisala com Provide at ...

Page 98: ...www vaisala com ...

Reviews: