Installation with Chemistry-HYBRID pump RC 6
+
Remove the hose nozzle (outlet) from the distributor plate (1).
+
Using scissors, cut the connecting tube (2) to a length of 18.5 cm.
+
Screw the threaded adapter (3) together with the seal ring (4) into the
distributor plate. Then screw the short elbow fitting (5) into the thread
-
ed adapter.
+
Assemble the connecting tube (2) to the exhaust waste vapor con-
denser. Slip the union nut onto the connecting tube (with backside
ahead, see (6)). Slip the connecting tube as far as it will go onto the
elbow fitting (5). Tighten the union nut (6) first by hand and then tighten
one full turn using an open end wrench.