![vacuumbrand PC 8 / MD 12C Instructions For Use Manual Download Page 29](http://html1.mh-extra.com/html/vacuumbrand/pc-8-md-12c/pc-8-md-12c_instructions-for-use-manual_3132834029.webp)
➨
Assemble the separator at the inlet using the
centring ring and the clamping ring.
Exhaust waste vapor condenser:
➨
Assemble the hose nozzles for coolant inlet
(13) and coolant outlet (14) tubing at the ex-
haust waste vapor condenser.
➨
Assemble the collecting flask at the exhaust
waste vapor condenser using a joint clip.
The
exhaust waste vapor condenser
enables an effi
-
cient condensation of the pumped vapors at the outlet.
+
No backflow of condensates.
+
Controlled recovery of condensates.
+
Close to 100% solvent recovery.
+
The isolation cover protects against glass splinters in
case of breakage, acts as thermal isolation to avoid
condensation of humidity and is intended to absorb
shocks.
➨
Attach the tubing of the coolant circuit to the respective
hose nozzles (hose nozzles for tubing I.D. 1/4”-5/16”
(6-8 mm), see image) at the waste vapor condenser.
Check all hose connections prior to starting operation of
the cooling system.
Secure coolant hoses at the hose nozzles (e.g., with hose
clamps) to prevent their accidentally slipping off.
➨
Prevent the discharge of dangerous gases and vapors
to the surrounding atmosphere. If appropriate, connect
the exhaust line to a suitable treatment system.
+
Never block the gas outlet (hose nozzle for tubing I.D.
3/8” (10 mm)). The exhaust hose has always to be un-
obstructed and without back pressure to enable an un-
hindered discharge of gases and protect the pump from
damage.
13
14
NOTICE