Continuous pumping with pressure and time settings
• Operation of a vacuum pump via isolation valve
• Operation of a vacuum pump without isolation valve with VMS (Vacuum Management System,
see ”Accessories”)
Preselections
☞
The selection of the parameters is carried out via the selection knob. All parameters are also change-
able during running operation control.
☞
Minimum:
The controller switches off the pump or closes the isolation valve when the preset value
for ”
Minimum
” is reached.
”
Minimum
” is adjustable in a range of 1-1060 mbar and to ”Off”.
A preset ”
Duration
” (process time) is without function, if the process is stopped due to a preset ”
Mini-
mum
” before ”
Duration
” is reached.
☞
Delay:
”
Delay
” determines the time the pump (with VMS and isolation valve) and the coolant valve
run on after stopping the process.
The ”
Delay
” is adjustable in a range of 1-300 minutes and to ”Off”.
☞
Duration:
”
Duration
” determines the total process time since its start.
The process time is adjustable between 1-1440 minutes (24 h) and to ”Off”. ”Off” indicates that no
endpoint for pump down is determined.
If a ”
Duration
” is preset, the controller switches off the pump if the preset process time is reached,
also if a preset ”
Minimum
” is still not reached.
☞
If neither ”
Minimum
” nor ”
Duration
” are preset, process control has to be stopped by pressing key
”STOP”.
The screen-shot shows the factory-set values.
Pump down
Minimum
Off
Delay
Off
Duration
Off
- - - - - - Graphic - - - - - -
- - - - - - Back- - - - - - - -
00:00:00
When selecting ”
Graphic
” the display shows a pressure over time curve.
The timeline in the diagram adapts automatically to the process time.
☞
Press the selection knob twice to return to the standard display.
Pump down
10 mbar
00:03:50
Temporary switching from ”Pump down” to ”Vac control” (only if control is running):
☞
Press key ”MODE”. The controller switches to function
”Vac control”
, the actual vacuum is used as
set value.
☞
The preset function of the controller does not change due to this temporary switching. When pressing
key ”STOP” the controller is in function
”Pump down”
again.
Function Pump down