Notes regarding the assembling of the diaphragm

The set of seals is designated for diaphragm pumps with different technical performance.

Delivering pumps with square hole diaphragm starts in the second term of 2002, but not at the same time
for all types of pumps. 

(Exception MZ 2D: Always diaphragm with round hole.)

Replacing a 

diaphragm with round hole

, use the delivered diaphragm completely.

Replacing a 

diaphragm with square hole

, remove the punched part in the centre of the diaphragm before

assembling.

Replacing a diaphragm with round hole

Check for washers. Do not mix the washers from the differ-
ent heads. Make sure that the original number is reassem-
bled at the individual pump head.

Smaller number of washers: The pump will not attain ulti-
mate vacuum. More washers: Clamping disc will hit head
cover; noise or even blockade of the pump.

Insert diaphragm support disc and new diaphragm.

Note: 

Double diaphragm.

Insert diaphragm clamping disc, apply a drop of adhesive
(OmniFit

®

 50M or Loctite

®

 243) to lowerside of screw head

and thread and screw diaphragm clamping disc, diaphragm
and diaphragm support disc to connecting rod.

Use Phillips screw driver, remove countersunk head screw
in the centre of the diaphragm clamping disc.

Note: 

Screw is secured by adhesive to prevent it from being

detached. Clean screw or use new one (cat. no.: 639847).

If the old diaphragm is difficult to separate from the support
disc, immerse assembly in naphtha or petroleum ether. Do
not inhale!

Check diaphragm for damage and replace if necessary.

To do so use Phillips screw driver to remove four counter-
sunk head screws and lift off housing plate with rating plate.

Use petroleum ether or industrial solvent to remove depos-
its if necessary.  Do not inhale.

OmniFit

®

 and Loctite

®

  are registered trade marks of Henkel Technologies

Summary of Contents for ME 2

Page 1: ...completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product manual no 995652 08 10 2009 Diaphragm pumps ME 2 ME 4 MZ...

Page 2: ...r pump VACUUBRAND diaphragm pumps are the result of many years of experience in construction and practical operation of these pumps combined with the latest results in material and manufacturing techn...

Page 3: ...ct to his product manual no 995652 08 10 2009 Attention Important notes Not permitted Misuse may cause damage Caution Hot surface Isolate equipment from mains Note Contents Safety information 4 Techni...

Page 4: ...protected electrical supply and a suitable earth point Failure to connect the motor to ground may result in deadly electrical shock The supply cable may be fitted with a moulded European IEC plug or...

Page 5: ...pheres If pumping different substances purge the pump with inert gas prior to changing the pumped media in order to pump out residues and to avoid reactions of the pumped substances with each other wi...

Page 6: ...ts The A weighed emission sound pressure level of the pump does not exceed 70 dB A Measurement according to EN ISO 2151 2004 and EN ISO 3744 1995 with standard silencer or exhaust tube at outlet Ensur...

Page 7: ...o m t a o n n o i t a r e p o g n i r u d t a s n e d n o c f o n o i t a r a p e s r e w o p r o t o M W 0 2 1 0 8 1 0 8 1 0 8 1 n o i t c e t o r p r o t o M t u o t u c l a m r e h t w a r d t n e...

Page 8: ...M P F 2 Z M 2 E M e l z z o n e s o h T B P 4 E M e l z z o n e s o h l e e t s s s e l n i a t s D 2 Z M e g n a l f l l a m s l e e t s s s e l n i a t s r e c n e l i s E P A P e s o h E P g n i t...

Page 9: ...of this document with respect to his product manual no 995652 08 10 2009 ME 2 with dual voltage motor 100 120 200 230V ME 2 outlet with silencer mains cable on off switch outlet with silencer inlet p...

Page 10: ...ct manual no 995652 08 10 2009 ME 4 MZ 2 MZ 2D fig MZ 2D inlet outlet with silencer pump rating plate motor rating plate on off switch mains cable inlet MZ 2 hose nozzle outlet pump rating plate motor...

Page 11: ...igid pipelines Insert elastic hoses or flexible elements as couplings between the pump and rigid pipes Use a suitable valve to isolate the pump from the vacuum system if you need to allow the pump to...

Page 12: ...oling down Take suitable precautions if an automatic restart of the pump may lead to a critical dangerous situation Due to the high compression ratio of the pumps pressure at the outlet port might be...

Page 13: ...roduct manual no 995652 08 10 2009 ME 2 MZ 2 Small flange DN 16 662590 Clamping ring DN 10 16 aluminium 660000 Centring ring DN 10 16 NBR 660190 Hose nozzle for MZ 2D 639758 ME 4 Small flange DN 16 63...

Page 14: ...840 Pressure adjustment device outlet 696841 Installation vacuum and or pressure adjustment device at ME 2 MZ 2 Assemble control valve first then align and fasten gauge Installation vacuum adjustment...

Page 15: ...a t l o v e v e i h c a t o n s e o d p m u P r o e r u s s e r p l a t o t e t a m i t l u d e e p s g n i p m u p l a m r o n y l t c e r r o c t o n g n i r g n i r e t n e C e h t n i k a e l r o...

Page 16: ...ump is exposed to corrosive gases or vapour or in case of deposits mainte nance should be carried out frequently Regular maintenance will improve the lifetime of the pump and also protect both man and...

Page 17: ...ound hole clamping disc with countersunk head screw diaphragm with square hole clamping disc with square head screw Use open ended wrench remove the screw in fittings together with the hose from all p...

Page 18: ...mp head Smaller number of washers The pump will not attain ulti mate vacuum More washers Clamping disc will hit head cover noise or even blockade of the pump Insert diaphragm support disc and new diap...

Page 19: ...eccentric bushing bring connecting rod into up per turning point position Max stroke of the rod Pay attention that the diaphragm stays positioned centrally so that it will become clamped uniformly bet...

Page 20: ...p chamber By measuring the pressure at the inlet port of the individual head Use a suitable vacuum gauge e g DVR 2 cat no 68 29 02 make sure that it is correctly calibrated and measure the pres sure a...

Page 21: ...ry product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair or from the pur ch...

Page 22: ...for the returned product that neither toxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oi...

Page 23: ...e dispositif d sign ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne conception dessin et mod le vendu par nous m mes Cette d claration ces...

Page 24: ...ature is only intended to inform our customer The validity of general empirical values and results obtained under test conditions for specific applications depend on a number of factors beyond our con...

Reviews: