Attention:
Before starting the pump, check that the voltage changeover switch at the
terminal box is correctly positioned. Change the selection at the voltage
changeover switch only if the pump is separated from mains.

Voltage changeover switch:

Use a screw driver to adjust the supply voltage on the 

voltage changeover switch

 at

the terminal box of the pump:
”115” corresponds to 100-120V and
”230” corresponds to 200-230V.

Use and operation

Installing in a vacuum system:

Avoid throttling losses by using connecting pipes with large diameter and keep
them as short as possible.

Reduce the transmission of vibration and prevent loading due to rigid pipelines.
Insert elastic hoses or flexible elements as couplings between the pump and rigid
pipes.

Use a suitable valve to isolate the pump from the vacuum system if you need to
allow the pump to warm up before you pump condensable vapours or if you clean
the pump before it is switched off.

Connect the exhaust to a suitable treatment plant to prevent the discharge of dan-
gerous gases and vapours to the surrounding atmosphere. Use a catchpot to pre-
vent the drainage of contaminated condensate back into the pump.

Before switching on the pump check that mains voltage and current conform
with the equipment!

Pump with dual-voltage motor:

Check that the voltage changeover switch at the terminal box is positioned
correctly.

Attention: If the pump is switched on with the voltage not correctly selected,
the motor may be damaged! 

Pay attention to the warning label on the on/off

switch (ME 2 only) and remove after having set the voltage changeover switch to
the correct position!

Selecting  the supply voltage:

ME 2

 with dual-voltage motor

MZ 2D

 with dual voltage motor

Summary of Contents for ME 2

Page 1: ...completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product manual no 995652 08 10 2009 Diaphragm pumps ME 2 ME 4 MZ...

Page 2: ...r pump VACUUBRAND diaphragm pumps are the result of many years of experience in construction and practical operation of these pumps combined with the latest results in material and manufacturing techn...

Page 3: ...ct to his product manual no 995652 08 10 2009 Attention Important notes Not permitted Misuse may cause damage Caution Hot surface Isolate equipment from mains Note Contents Safety information 4 Techni...

Page 4: ...protected electrical supply and a suitable earth point Failure to connect the motor to ground may result in deadly electrical shock The supply cable may be fitted with a moulded European IEC plug or...

Page 5: ...pheres If pumping different substances purge the pump with inert gas prior to changing the pumped media in order to pump out residues and to avoid reactions of the pumped substances with each other wi...

Page 6: ...ts The A weighed emission sound pressure level of the pump does not exceed 70 dB A Measurement according to EN ISO 2151 2004 and EN ISO 3744 1995 with standard silencer or exhaust tube at outlet Ensur...

Page 7: ...o m t a o n n o i t a r e p o g n i r u d t a s n e d n o c f o n o i t a r a p e s r e w o p r o t o M W 0 2 1 0 8 1 0 8 1 0 8 1 n o i t c e t o r p r o t o M t u o t u c l a m r e h t w a r d t n e...

Page 8: ...M P F 2 Z M 2 E M e l z z o n e s o h T B P 4 E M e l z z o n e s o h l e e t s s s e l n i a t s D 2 Z M e g n a l f l l a m s l e e t s s s e l n i a t s r e c n e l i s E P A P e s o h E P g n i t...

Page 9: ...of this document with respect to his product manual no 995652 08 10 2009 ME 2 with dual voltage motor 100 120 200 230V ME 2 outlet with silencer mains cable on off switch outlet with silencer inlet p...

Page 10: ...ct manual no 995652 08 10 2009 ME 4 MZ 2 MZ 2D fig MZ 2D inlet outlet with silencer pump rating plate motor rating plate on off switch mains cable inlet MZ 2 hose nozzle outlet pump rating plate motor...

Page 11: ...igid pipelines Insert elastic hoses or flexible elements as couplings between the pump and rigid pipes Use a suitable valve to isolate the pump from the vacuum system if you need to allow the pump to...

Page 12: ...oling down Take suitable precautions if an automatic restart of the pump may lead to a critical dangerous situation Due to the high compression ratio of the pumps pressure at the outlet port might be...

Page 13: ...roduct manual no 995652 08 10 2009 ME 2 MZ 2 Small flange DN 16 662590 Clamping ring DN 10 16 aluminium 660000 Centring ring DN 10 16 NBR 660190 Hose nozzle for MZ 2D 639758 ME 4 Small flange DN 16 63...

Page 14: ...840 Pressure adjustment device outlet 696841 Installation vacuum and or pressure adjustment device at ME 2 MZ 2 Assemble control valve first then align and fasten gauge Installation vacuum adjustment...

Page 15: ...a t l o v e v e i h c a t o n s e o d p m u P r o e r u s s e r p l a t o t e t a m i t l u d e e p s g n i p m u p l a m r o n y l t c e r r o c t o n g n i r g n i r e t n e C e h t n i k a e l r o...

Page 16: ...ump is exposed to corrosive gases or vapour or in case of deposits mainte nance should be carried out frequently Regular maintenance will improve the lifetime of the pump and also protect both man and...

Page 17: ...ound hole clamping disc with countersunk head screw diaphragm with square hole clamping disc with square head screw Use open ended wrench remove the screw in fittings together with the hose from all p...

Page 18: ...mp head Smaller number of washers The pump will not attain ulti mate vacuum More washers Clamping disc will hit head cover noise or even blockade of the pump Insert diaphragm support disc and new diap...

Page 19: ...eccentric bushing bring connecting rod into up per turning point position Max stroke of the rod Pay attention that the diaphragm stays positioned centrally so that it will become clamped uniformly bet...

Page 20: ...p chamber By measuring the pressure at the inlet port of the individual head Use a suitable vacuum gauge e g DVR 2 cat no 68 29 02 make sure that it is correctly calibrated and measure the pres sure a...

Page 21: ...ry product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair or from the pur ch...

Page 22: ...for the returned product that neither toxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oi...

Page 23: ...e dispositif d sign ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne conception dessin et mod le vendu par nous m mes Cette d claration ces...

Page 24: ...ature is only intended to inform our customer The validity of general empirical values and results obtained under test conditions for specific applications depend on a number of factors beyond our con...

Reviews: