page 65 of 75
holder of the exhaust vapor condenser with
a Torx driver T10. Remove the exhaust va-
por condenser. In doing so pull the PTFE
hose out of the inlet of the exhaust vapor
condenser.
If necessary, pull off the old pressure relief
valve and install the new one. Check for
correct position of the PTFE-foil under the
pressure relief valve.
Feed the PTFE hose into the inlet of the
exhaust vapor condenser. Mount the con-
denser with the counter holder (Torx screws
T10) to the pump. Fasten union nut.
Overpressure
safety relief device
............................
638821
(at exhaust waste vapor condenser)
Remove catchpot (9) at outlet (see ”Use
and operation”, pg. 38).
-
haust vapor condenser.
Replacing the overpressure safety relief device