vapors at the outlet.
Controlled recovery of condensates.
Close to 100% solvent recovery.
The isolation cover protects against glass
splinters in case of breakage, acts as ther-
mal isolation to avoid condensation of hu-
midity and is intended to absorb shocks.
Attach the tubing of the coolant circuit to the respective
hose nozzles (hose nozzles for tubing I.D. 1/4”-5/16”
(6-8 mm), see image) at the waste vapor condenser.
Check all hose connections prior to starting operation of
the cooling system.
Secure coolant hoses at the hose nozzles (e.g., with hose
clamps) to prevent their accidentally slipping off.
Prevent the discharge of dangerous gases and vapors
to the surrounding atmosphere. If appropriate, connect
the exhaust line to a suitable treatment system.
Never block the gas outlet ((2) hose nozzle for tubing
I.D. 3/8” (10 mm)). The exhaust hose has always to be
unobstructed and without back pressure to enable an
unhindered discharge of gases and protect the pump
valves from damage.
Note
: Install the hoses of the cooling system in a way
pumping unit (especially cables and electronic parts,
see also IP degree of protection, ”Technical data”, pg.
24).
coolant outlet tubing
is always unob-
structed and that it cannot get blocked.
waste vapor condenser: 87 psi (6 bar) absolute. Outlet
-
sures of additional components in the coolant circuit
NOTICE
Summary of Contents for MD 4C NT + 2 AK
Page 25: ...page 30 of 75 ME 4C NT 1 5 4 7 6 8 2 ME 2C NT 1 5 4 7 6 8 2 ...
Page 27: ...page 32 of 75 ME 8C NT MD 4C NT 1 5 3 4 7 6 2 8 8 5 4 7 6 1 ...
Page 67: ...page 72 of 75 tive mark Licensed Test mark on the pump rating plate ...
Page 68: ...page 73 of 75 China RoHS ...
Page 69: ...page 74 of 75 ...