Level limit switch
Series VN 1000 / 2000 / 5000 / 6000
Technical information / Instruction manual
page
Changing the Electronic module
Intrinsic safe marked Electronic modules are not allowed to be exchanged with
Electronic modules without Intrinsic safe marking. Observe warning labels
inside the housing and Ex marking on the name plate.
1.
Open the housing lid, remove the pigtails from the device.
2.
Disconnect internal wire for earth connection (not on all versions).
3.
Unscrew two fastening screws of the electronic module.
4.
Pull out the Electronic module.
5.
Insert a new Electronic module (until it locks into place) and tighten fastening screws.
6.
Connect internal wire for earth connection (not on all versions).
7.
Connect the pigtails to the device.
Opening the lid (cover)
Before opening the lid for maintenance reasons observe following items:
• Do not remove the lid while circuits are alive.
• No dust deposits or whirlings are present.
• No rain can enter into the housing.
Frequent check of the
unit
Frequent check of the unit
To ensure durable safety in hazardous locations and with electrical safety, following items must be
checked frequently depending on the application:
• Mechanical damage or corrosion of any components (housing side and sensor side) and of the field
wiring cables.
• Thight sealing of the process connection, cable glands and enclosure lid.
• Properly connected external PE cable (if present).
Cleaning
Cleaning
If cleaning is required by the application, following must be observed:
• Cleaning agent must comply with the materials of the unit (chemical resistance). Mainly the lid
sealing, cable gland and the surface of the unit must be considered.
The cleaning process must be done in a way, that:
• The cleaning agent cannot enter into the unit through the lid sealing or cable gland.
• No mechanical damage of the lid sealing, cable gland or other parts can happen.
A possible accumulation of dust on the unit does not increase the maximum surface temperature and
must therefore not be removed for purposes of maintaining the surface temperature in hazardous
locations.
Function test
A frequent function test may be required depending on the application.
Observe all relevant safety precautions related with a safe work depending on the application (e.g.
hazardous locations, hazardous bulk material, electric safety, process pressure).
This test does not proof if the sensor is sensitive enough to measure the material of the application.
Function test is done by stopping the vibration of the vibrating rods with appropriate means and
monitor if a correct change of the signal output from uncovered to covered happens.
Production date
The production date can be traced by the serial number on the typeplate. Please contact the
manufacturer or your local distrubutor.
Spare parts
All available spare parts are stated in the selection list
Maintenance
Repair of flamepath
Repair of flamepath on units with Ex d, Ex de or XP approvals is not intended.
Please contact manufacturer.