background image

 

Intended use: The WPT-900 portable transmitter 

utilizes 900MHz Frequency Agile® technology.  

 

The WPT-900 has a water-resistant housing that 

protects the pendant when worn in the shower, rain, 

or other environments where the pendant may be 

splashed with water. It is not, however, designed to 

be submerged

 
 
 
 
 
 

 

 

               

 

 
 
 
 

 

 
 

 

 

 

WPT-900 

Installation

Instructions

To activate the WPT-900, simply step out of the range radius of receiver unit.  
 
Pendant Check-in signals are transmitted every 10 seconds, and are indicated by the blinking 
transmit LED (the little light in the upper-left corner of pendant). When the WPT-900 comes back in 
range of receiver unit, it will begin to check in with the receiver unit once again, and the alarm 
condition will end. 
 
Changing the battery: 1) Slip off clip from back of pendant 2) Remove 3 screws from the back of the 
housing 3) Make sure that the battery is inserted with the positive (+) terminal facing up 4) After 
replacing the battery, press the internal reset button to reset the transmitter to its most recent 
programming 5) Replace back cover and carefully re-tighten screws evenly for tight seal 6) 
Thoroughly test the unit before returning to service. 
 

NOTE: 

WPT-900 transmitters retain programming data in non-volatile memory. They do not 

require reprogramming after loss of power. Install new battery and press either the internal or the 
external reset button to re-initialize the transmitter and restore programming.  
  

Pendants should be tested at least once per week. 

 
The WPT-900 can be carried in 4 convenient ways 

 

Around the neck as a pendant (standard). 

 

On a belt with the belt clip attachment (standard). 

 

On the arm with non-removable locking wristbands.  

 

On the ankle with a non-removable locking bands. 

 
WARNING:  The neck cord provided with the WPT-900 is designed for user safety. Substitution of 
stronger cords or chains may result in injury to the wearer. 

Dimensions: 2.22” x 1.925” x .725”  Weight:  1.4 oz. 
Battery:  3.0v Lithium Sanyo CR2450 or Renata CR2450N 
Typical Battery Life: 1 to 2 years (Battery is always supervised. The receiver will 
notify you when battery is running low). 

Summary of Contents for Wanderer Alert WA-900

Page 1: ...ANDERER ALERTTM The pendant Transmitter automatically sends a signal every seconds to the receiver dialer If the receiver doesn t hear from the pendant within an adjustable period indicating patient h...

Page 2: ...ment off then on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate radio or television Increase the separation between the equipme...

Page 3: ...ntation WA 900 Pendant Movements WA 900 placed On Near metal surfaces Cross traffic Furniture Interior and Exterior Walls Building Structures Holding the transmitter next to your body or the ground wh...

Page 4: ...system and Momentary Continuous activation Special Features The dialer will beep every 30 seconds in the Program and Test mode This is to warn the user that the dialer is not in the Operate or Off mod...

Page 5: ...e illness sufferers and others who desire the reassurance of 24 hour peace of mind Combined with the cordless WPT 900 Transmitter The WA 900 will call for help even if you are unable to reach the tele...

Page 6: ...g telephone line cord 3 WPT 900 Cordless Transmitter 4 14V AC DC adaptor with 3 5mm plug 5 WPT 900 Transmitter Locking Band Not Shown 6 Wanderer Alert Unit 7 Instructional Pack Quick Start Guide Range...

Page 7: ...ery press either the internal or the external reset button to re initialize the transmitter and restore programming either the small black button located on the back of the outside of the pendant or t...

Page 8: ...VOICE VERIFICATION OGM AND 2 WAY SPEAKER FOR PLAYBACK MODE PROGRAM RECORD VOICE PAUSE DIGIT 2S DIGIT OPERATE OFF TEST MICROPHONE TO RECORD VOICE MESSAGE EASY TOUCH KEYPAD EASY PROGRAMMING WITH M 0 P...

Page 9: ...I n s t a l l a t i o n H o o k u p 8 Important This configuration should only be setup by a professional technician that has electrical wiring experience...

Page 10: ...ss the internal reset button to reset the transmitter to its most recent programming 5 Replace back cover and carefully re tighten screws evenly for tight seal 6 Thoroughly test the unit before return...

Page 11: ...be input in each phone pager location The P Pause Button Press this key to program in a pause Each P provides a 2 second pause Pauses are mainly used to compensate for connection lags such as when ca...

Page 12: ...plained in step 3B cellular phone numer explained in step 3B or a voice pager explained in step 3C VERY IMPORTANT When programming in the first number whether it is a pager or not it s necessary to pr...

Page 13: ...numbers first press 1 followed by the area code and then the number For CELLPHONES enter 3 pauses press P button 3 times after the digits of the phone number but before pressing the asterisk to recor...

Page 14: ...get an outside line with the PBX feature activated or if your phone system requires more than one digit to get an outside line it will be necessary to disable the PBX feature and instead program in t...

Page 15: ...2 NO B Press 1 to program your OGM Press 2 if you have already recorded your voice message or if you are only calling numeric pagers that don t require a voice message STEP 3 After selecting your opt...

Page 16: ...ther because of an out of range transmitter or a trigger on the hard wired input it will remain in alarm until the unit calls all programmed phone numbers or the unit is manually reset even if the tra...

Page 17: ...vered Press 2 to scroll to the next section STEP 5 To test the operation of your entire system set Mode to OPERATE and proceed as explained in the next section Operating Your System NOTE IN ORDER TO T...

Page 18: ...or High volume setting All programming information will be retained Input Activation The receiver dialer goes into alarm when either the check in signals from the pendant transmitter are not received...

Page 19: ...er dialer into the OFF mode and move the volume switch on the horn strobe to the off position Check in signals are transmitted every 10 seconds and are indicated by the blinking transmit LED When the...

Page 20: ...ROGRAM OGM prompt STEP 7 Using a prepared script and speaking six to eight inches from the dialer s microphone on the front of the unit press R when you are ready to enunciate your emergency message S...

Page 21: ...__ Week 34 ___________ Week 8 ___________ Week 35 ___________ Week 9 ___________ Week 36 ___________ Week 10 ___________ Week 37 ___________ Week 11 ___________ Week 38 ___________ Week 12 ___________...

Page 22: ...ge without notice Dialer Accessories S p e c i f i c a t i o n s Power Source PB12P AC DC Adaptor With Power Pack Rechargeable Plugs into regular 110VAC outlet to provide primary and backup power AC 2...

Reviews: