UIT2400 INSTRUMENTATION MANUAL REV A, May 2020 

7 OF 18 

4B. Charging: 

 Charging the battery is done through the USB 3.0 mini or micro port. 

 

 

WARNING: The battery pack in the UIT2400 is expected to last for many years. (up to 500 charge 

cycles).  Batteries should only be charged, stored and used in temperature between 0-45 degrees C, 

or expansion and damage can occur.  Battery packs should only be replaced by trained personnel. 

Changing of the battery pack by non ILT personnel voids the warranty and can cause damage to the 

pc board.

 

 

The power source can be supplied from a computer, a power bank, a charging hub, a wall-wart or any 

USB 2.0 or 3.0 device. A full charge is obtained in approximately 12 - 14 hours.  Meter will charges the 

fastest when the power button is in the off position. The UIT2400 circuit is design to prevent over 

charge. The UIT2400 has a low power indicator that will advise when the battery needs recharging. 

 

 

5. Sensor Connector and Wiring:

 

The Smart sensors are supplied with an internal memory chip inside the D-sub connector that 

includes the model and serial number of the sensor as well as calibration information. 

Plug in 

the sensor before turning on the UIT2400 to enable smart sensor functionality.

 

 

The D-sub not only receives the current generated by the sensor after exposure to light, it also 

supplies bias to sensors requiring bias for power and/or to improve response time.  

 

6. USB Ports: 

There are three USB ports on the side of the UIT2400 device

 

 

UPDATE:

 Is used for code updates and requires the use of a memory stick. (not 

included) 

 

 

DATA:

  Accesses the internal 8GB memory device where data is saved and to charge 

the battery. The Micro USB is used to download stored data and to erase data from the 

memory device.  This port can also be used the charge the battery.  

 

 

PC:

  The Mini USB port is used to connect to a computer and to charge the battery.   

Datalight software is available for download on the website: https://www.intl-

lighttech.com/support/software, or request an emailed copy by emailing ilsales@intl-

lighttech.com

 

 

7. Display:

 

The Display is a full color

 4.3" Capacitive TFT Touch screen display.  Contact with sharp 

objects will damage the display. A stylus or finger is recommended for changing settings on the 

display. Care should be taken when packaging and storing the meter to assure no pressure is 

applied to the surface of the display.

 

Do not expose the Heat Sources, Strong UV Sources, 

liquids or Chemicals.

 

 

8. Current Measurement and Limitations:

 

The UIT2400 provides accurate/calibrated measurements in the 50 pico-amp to 1 milli-amp 

range and can be manually fixed to a specific range, or automatically change to the correct 

range as the light level changes, using the auto range feature.

 

 

8A

Saturation

: The UIT2400 provides an autorange mechanism allowing it to measure over 8 

decades of current and/or light levels. While autoranging the UIT2400 can experience 

saturation. When this condition occurs the display will provide a “SATURATION” warning. To 

prevent saturation, neutral density filters or apertures can be added to the sensor, or a greater 

Summary of Contents for UIT2400

Page 1: ...USHIO AMERICA INC 5440 Cerritos Avenue Cypress CA 90630 Tel 714 236 8600 Fax 714 229 3180 www ushio com UIT2400 Operation Manual Ushio America Inc...

Page 2: ...Wiring 7 6 USB Ports 7 7 Display 7 8 Current Measurement and Limitations 7 9 Sample Time and Limitations 8 10 5V Bias for silicon sensors 8 11 Home Screen 8 12 Display Style Selection 8 13 Display Co...

Page 3: ...ata retrieval C SED220 custom 222nm detector NIST traceable calibrated at 222nm D 8GB flash memory for measurement recordings E CC200 Case Protective carrying shipping case F Electrical Calibration Ce...

Page 4: ...charging 3D Top Contains the sensor mount a 15 Pin D Sub connector and posts for securing the sensor to the device To connect the sensor align the sensor s male 15 pin connector to the meter s female...

Page 5: ...and a scroll menu will appear select a new channel until you find the correct sensitivity factor as noted on your certificate If you do not locate your calibration factor and units from the certificat...

Page 6: ...ure readings see 3F2 above to learn more about assuring an accurate zero 3G7 Hold Freezes the numerical display value on screen 3G8 100 Takes a measurement and saves it as the 100 baseline for transmi...

Page 7: ...n the side of the UIT2400 device UPDATE Is used for code updates and requires the use of a memory stick not included DATA Accesses the internal 8GB memory device where data is saved and to charge the...

Page 8: ...test settings Note There are multiple factors that affect the sampling time including the sensors rise time the light level software selection software settings and the method used to store and transf...

Page 9: ...cases the level can be used as a go no go level indicating the risk of product failures when intensities are too low or too high YELLOW is typically used as a warning stage indicating the range has c...

Page 10: ...ver time When in integrate mode the units of the readings change to include seconds For example W cm2 becomes W cm2 seconds or J cm2 W m2 becomes J m2 14D Hold Freezes the value on the numerical displ...

Page 11: ...urrent 1uA 0 5sec is used else if current 1uA 1 0sec FAST samples at 20ms 50Hz Note Using trend at fast speeds taxes the CPU with processing both measurements numerical updates and graph updates This...

Page 12: ...Interval Ranges from 1 second to 1 hour or sample for faster speeds Sample when selected the record interval and sample time are synched Ie if sample time is set 1 the meter will record a value every...

Page 13: ...vice turns on 15M Use Scientific Notation ON OFF Scientific notation ON uses an exponent to show the decimal place when off a prefix is used Examples On 1 35 e 3 W cm2 Off 1 35 mW cm2 On 4 78e 6 W cm2...

Page 14: ...calibration certificate 2 09e 09 see red arrow on cal cert on page 14 15Q 5 Resulting Light Level Allows the user to enter a known value for the light level under test and create a calibration factor...

Page 15: ...ouble click on the Removable Disk within Devices with Removable Storage To open in excel you will need to select tab or space delimited 16D Deleting Stored Readings To delete files access the files th...

Page 16: ...elect Location box 9 To save reading s touch the capture button Readings will be saved to the Survey_data_file csv file accessible fie the usb micro port 20 Info Lists product and software settings in...

Page 17: ...f the meter Always store the meter in the carrying case when not in use and always ship in the hard plastic carrying case with protective packing to prevent shock and damage No items should be place a...

Page 18: ...America customer service and obtain an RMA customerservice ushio com 24 Warranty The equipment you have purchased from Ushio America Inc has been expertly designed and was carefully tested and inspect...