IM 50.560CA UA (B/4-03)
15
DEPOLOX
®
3
PLUS
RESIDUAL ANALYZER
2.6
Flow Switch (AAA7000) (Optional)
In general, the flow switch is only connected to the digital input of the
DEPOLOX 3 plus. In case of other applications, ensure that the rating of
this Reed contact (100 V AC and 0.5 A) is not exceeded even for a short
time. In case inductive loads (e.g., relays or contactors) are switched, the
contact has to be protected against induction sparks.
Figure 2.3 - Flow Switch
See Dwgs. 50.560.100.020 and .030 for flow switch location.
2.7
Stop Valve (U95687) (Optional)
Figure 2.4 - Stop Valve
See Dwgs. 50.560.100.020 and .030 for stop valve location.