14
5 | Maintenance
Shut off the power switch AND disconnect the power cord before any maintenance.
Use a damp soft cloth to wipe clean. Stubborn stains should be cleaned by neutral detergents.
The maintenance of internal electrical and heating parts must be performed by professionals or trained
electricians.
Do not directly splash water over the product or use abrasive powder, diluent, oil, kerosene, acidic material,
and similar substances during cleaning, or else shock or other accidents will occur.
5.1 Periodic Maintenance
Fluid:
If using Silicone Oil, degradation is not much of a concern. Keep an eye on the tank level.
Propylene Glycol will degrade over time, it is recommended to purchase glycol test strips.
Unit:
Check the unit over monthly for leaks, broken glass, melting, burning, or any other damages.
Please bring any concerns to our attention immediately.
5.2 Long Term Storage
Empty the unit of fluid.
Cap the inlet and outlet.
Clean the condenser.
Clean the body of the unit.
Wrap the unit with plastic.
Store in a secure level location.