EXAMET-5 SERIES MICROSCOPE
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com
9
DETAILED ASSEMBLY
(continued)
Installing the Reflected Illuminator
(Fig. 4)
Completely loosen the lock screw
①
of the illuminator.
As shown, position the illuminator above the dovetail hole, tilt
it slightly down on the left and carefully slide the dovetail
under the notches in the dovetail hole and set it into place.
Make the sure it is seated properly, then tighten the lock
screw
①
.
Installing the Head
(Fig. 5)
Completely loosen the lock screw
①
of the head.
As shown, position the head and with it tilted slightly down on
the right, slide the dovetail under the notches in the dovetail
hole and set it into place with the two eyepiece tubes facing
forward. Tighten the lock screw
①
.
Installing the Objectives
(Fig. 6)
Rotate the coarse focusing knob to move the stage into its
lowest position.
Install the objectives
①
into the objective nosepiece
from
the lowest magnification to the highest in a clockwise
direction from the rear.
①
Fig. 4
Fig. 5
Fig. 6
①
①