EXAMET-5 SERIES MICROSCOPE
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com
19
TROUBLESHOOTING
Under certain conditions, performance of this unit may be adversely affected by factors other than
defects. If a problem occurs, please review the following list and take remedial action as needed. If you
cannot solve the problem after checking the entire list, please contact your local dealer for assistance.
OPTICAL SYSTEM
Problem
Cause
Solution
The bulb is bright
but it is dark in
field of view
Field diaphragm is not large enough
Largen the field diaphragm
Polarizer or analyzer is used
Pull them out
The beam splitter is in the trinocular
observation position
Move the beam splitter lever to the binocular
observation position
The side of the
field of view is
dark or not even
The nosepiece is not in the right position
Turn the nosepiece into the correct position
Stain or dust has accumulated on the lens
(objective or eyepieces)
Clean the lens
The color filter, polarizer or analyzer is not
in the correct position
Insert them or move out of the light path
Beam splitter is not in right position
Move it into the correct position
Stain or dust is
observed in the
field of view
Stains have accumulated on the specimen
Clean the specimen
Stains have accumulated on the lens
Clean the lens
Unclear image
Cover glass on the specimen slide
Use the one without cover
The specimen and the objective are not
vertical
Adjust it
The aperture is not opened correctly
Adjust it
Stain or dust has accumulated on the lens
of eyepiece
Clean the lens
Beam splitter is not in right position
Move it into the correct position
One side of the
image is dark or
the image moves
while focusing.
The specimen slide is not clamped
Clamp it with the stage clips
The nosepiece is not in the correct position Turn the nosepiece into the correct position
The eyes feel tired
easily -- the right
field of view
doesn’t
superimpose with
the left
Interpupillary distance is wrong
Adjust the interpupillary distance
Diopter adjustment is wrong
Adjust the diopter
The eyepiece for the right eye is different
from the left one
Use the same eyepieces