EXAMET-5 SERIES MICROSCOPE
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com
18
OPERATION
(continued)
Using a Microscopy Camera
(Fig. 31)
(Optional)
1. Loosen the lock screw
①
and remove the dust cap
②
on
the camera port
.
2. Remove the dust cover
③
on the c-mount adapter and
thread the adapter onto the camera.
3. Install the camera with the c-mount adapter into the photo
port.
4. Tighten the lock screw
①
.
5. Pull the light path selection lever
④
out to allow light to
enter the camera for observation.
Replacing the Fuse
(Fig. 32)
Turn the power switch to “O” (OFF) before replacing the
fuse.
Unplug the power cord from the outlet and disconnect it from
the back of the microscope.
NOTE
: there are two fuse holders on the back of the
microscope. The top fuse holder corresponds to the
top/reflected illuminator; the bottom fuse holder corresponds
to the bottom/transmitted illuminator.
Unscrew the fuse holder corresponding to the illuminator
that is not lighting with a flat head ( -- ) screwdriver.
Install a new fuse and re-install the fuse holder and tighten
with the screwdriver.
Plug the power cord into the back of the microscope and
into the wall outlet.
Turn on the microscope ( l ).
Use only the appropriate fuse for replacement: 250v 3.15a
(CAT# 142-30-03).
②
③
④
⑤
①
Fig. 31
①
Fig. 32