EXAMET-5 SERIES MICROSCOPE
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com
20
MECHANICAL SYSTEM
Problem
Cause
Solution
The objective
touches the cover
glass while
turning the
nosepiece
Stage is too high
Lower it to an appropriate position
Coarse focusing
knob is too tight
Tension adjustment ring is too tight
Loosen it to an appropriate position
Stage moves and
cannot stay on
the focal plane
Tension adjustment ring is too loose
Tighten it to an appropriate position
Coarse focusing
knob cannot rise
The coarse limit screw is locked
Loosen the screw
The image moves
obviously when
touching the
stage
The stage is fastened incorrectly
Fasten the stage correctly
ELECTRICAL SYSTEM
Problem
Cause
Solution
The bulb does not
work
No power supply
Check the connection of the power cord
The bulb burnt out
Replace it
The field of view
is not bright
enough
The light adjustment knob is not adjusted
correctly
Adjust correctly
The bulb flickers
or the brightness
is not stable
The bulb will burn out soon
Replace with a new one
The wire doesn’t connect well
Connect correctly