background image

P a g e

2

Rev. B

Proper Loading Instructions

1.

Plug the pump in. (The plug icon lights whenever the pump is plugged in. When the pump is

not plugged in and is battery-powered, the battery icon will light.)

2.

Prepare solution. Prime the set. Ensure all air is expelled from the set.

3.

Close set regulating clamp.

4.

Open pump door.  Raise pump door latch to horizontal position and pull door open.

5.

Load the set.

a.

Press

SAFETY CLAMP

to open position.

b. If your hospital has enabled the slide clamp loading feature

, insert the slide clamp on the

I.V. set fully into the slot, so that it is flush with the pump housing. (See

Illustration 1

.)  If the slide

clamp loading feature is not enabled, a black spring retainer occupies the slide clamp slot
and the slide clamp cannot be inserted.

c.

Load the tubing through the guide channel from bottom to top as shown in

Illustration 1

.

d.

Ensure that the tubing is loaded straight through the pump mechanism tubing guides

and safety clamp.

e.

Ensure that the tubing is touching the pumping fingers before closing the pump door.

6.

Close pump door. If resistance is felt when closing the door, check for misloaded I.V. set.

7.

Open set regulating clamp completely.

a.

Verify that no drops are falling in drip chamber. If flow is observed, close regulating

clamp, recheck I.V. set loading and verify that the proper administration set is being used.

b.

If flow is again observed, do not use the pump. Have it inspected by service personnel.

Note:

Always close the administration set regulating clamp(s) before opening pump door

and removing set.

8.

Attach set to I.V. access site.

9.

Turn pump on by pressing the

ON/OFF CHARGE

key. Verify that the pump performs  the

following self-test:

a.

All segments of both displays appear momentarily.

b.

Three separate audible tones sound.

c.

If a Hospital Area Designator (HAD) has been programmed into the pump, it appears for

3 seconds.

d.

The occlusion detection level is momentarily displayed in the message display (

LEVEL 1, 2,

or

3

), followed by

AUDIBLE SWITCHOVER

if the Audible Switchover option is enabled.

e.

If Auto Restart and Flow Check are both enabled, the message

AUTO RESTART

appears

for one second following the occlusion detection level display.

f.

If the pump is plugged into an AC outlet, the plug icon is lit. If the pump is running on

battery power, the battery icon is lit.

g.

If the message

INSERT SLIDE CLAMP

appears when you close the door, the slide clamp

loading option is enabled. Close the I.V. set's regulating clamp, open the door, and insert the
I.V. set's slide clamp fully into the slide clamp slot located below the safety clamp. Close the
door and open the regulating clamp.

10.

Set

VOLUME

knob on the rear of the pump to the desired level.

Summary of Contents for Baxter 6201

Page 1: ...e to tug tubing This will ensure an occlusion was not created when closing the door and will also make sure the tubing isn t being pinched at all by the clip Tip 3 To Avoid Nuisance Occlusion Alarms If you ve been using the tubing for a while try using a fresh section of the tubing by moving the clip down and loading this new section of tubing Tip 4 To Avoid Nuisance Occlusion Alarms Occasionally ...

Page 2: ...erved close regulating clamp recheck I V set loading and verify that the proper administration set is being used b If flow is again observed do not use the pump Have it inspected by service personnel Note Always close the administration set regulating clamp s before opening pump door and removing set 8 Attach set to I V access site 9 Turn pump on by pressing the ON OFF CHARGE key Verify that the p...

Page 3: ...set that error code quickly and easily with the process detailed on pages 5 6 of this document so you won t have to send in your pump In some cases the F49 or F94 error code is caused by a bad battery If after completing this process you believe that this is the case please contact us so we may help you verify this and if so we will assist you with returning your pump for correction You can reach ...

Page 4: ... can be adjusted to suit the needs of the hospital 4 Air sensor Detects air bubbles in the tubing The sensitivity is adjustable via the pump s configuration 5 Safety clamp Prevents accidental fluid flow by automatically occluding the tubing whenever the pump door is opened Provides Free Flow Protection 6 Slide clamp slot When the slide clamp loading option is enabled the I V set s slide clamp must...

Page 5: ...s Tot Vol Status and has the text Next lit above it 9 This key will be used to cycle through the screens of the Modify Config mode 10 Press this key approx 18 times until you see the top screen say Time with some numbers below it 11 Don t worry about what time the pump is currently saying we are now going to program in the current time at your location in 24 hour or Military format I E if it s 2 3...

Page 6: ... mode and be ready for use or it may come up with an additional Common Failure code of F 91 22 F 91 error is a common occurrence after this process and simply needs to have the pump powered off then back on again using the On Off key 23 Your pump should now be ready for use Although they can be caused by other factors typically these errors are caused by the main battery of the pump being complete...

Reviews: