P a g e
|
6
Rev. B
19.Once the date is inputted correctly in the bottom screen, press the “Pri Start” key and you
should see the date move from the bottom screen to the top screen and will display as:
11/19/11.
20.You are now done resetting the pump and can use the “On/Off” button to power down
the pump, then power it back up.
21.Upon power up you may see one of two things. It will either come up in the normal mode
and be ready for use or it may come up with an additional Common Failure code of F-91
22.F-91 error is a common occurrence after this process and simply needs to have the pump
powered off then back on again using the “On/Off” key.
23.Your pump should now be ready for use.
Although they can be caused by other factors, typically these errors are caused by the main
battery of the pump being completely drained. After completing this process the pump should be
allowed to charge for a solid 12 hours. If after a full charge this problem repeats, please contact
UnitedRx IV Department at 708-449-7600 x 335.