![United Process Controls Probe Buddy User Manual Download Page 9](http://html1.mh-extra.com/html/united-process-controls/probe-buddy/probe-buddy_user-manual_4009415009.webp)
PROBE BUDDY – USER MANUAL
Page 9 of 27
UPC_MAN013_R7
Copyright © United Process Controls.
All rights to copy, reproduce and transmit are reserved.
determines how long the unit will remain on, without any key being pressed, during the
specified period of time. This feature is available to minimize power drain on the battery.
Input Mode
The input function has been selected with the calculated oxygen
display for the thermocouple and millivolt input signals.
Pressing the Enter key cycles between the available displays
1.6.1 Using Input Mode
Input mode is used to check or test the outputs of a sensor in a working furnace. The sensor
should be at or above its minimum temperature range 1100°F (600°C). Disconnect the sensor
from the monitoring or control system and connect the PROBE BUDDY to the TC and mV
signal connections from the sensor.
If you are using the cable set with alligator clips, there will be a cold junction error introduced to
the temperature measurement, but the intention here is to confirm operation of the TC and
confirm the process value.
Output Mode
The output function has been selected with the calculated oxygen
display for the selected thermocouple and millivolt outputs.
Pressing the TC/mV key cycles between the TC/mV output
values. The
◄ or ►
keys move the cursor on the digits of the
value selected.
1600° F 1.12%
72.0 mV %O2
IMPEDANCE 1.20K
RECOVERY 10 SEC
→
TC TYPE - K
COLD JUNC 65
Enter
Enter
Enter
1600° F 1.12%
72.0 mV %O2
→