3. Connecting with indicators
15
Connecting through the SI/F
The M252A is equipped with a 2-ch SI/F. Parallel two-core cables or captyre cables are adequate for
use. The SI/F has no polarity.
1.
Strip the casing 0.2in (6mm) on the cable to be connected.
2.
Twist the bare wire to fit the terminal hole.
3.
Insert the supplied screwdriver into the lower hole and lightly hold it
down.
4.
Insert the twisted wires into the upper hole.
5.
Pull the screwdriver out from the lower hole.
6.
Make sure cable is clamped securely and does not come out with a
slight tug.
■
Sectional area of electric wires that can be connected to terminal blocks are 0.2 - 2.5mm
2
.
■
Do not attach crimp contacts to the ends of electric wires or solder them.
■
If you connect two or more electric wires, twist them beforehand.
■
Be aware that the orientation of the terminal block is upside-down to our digital indicators
(F366, etc.)
0.2in
Note