Decommissioning
Manual Pury 250e
20
Manual Pury 250e 2006-06-08
6.2.
Cooling block with fan
Clean the cooling body (6) as necessary for peak performance.
Figure 7 : Cooling block with finned cooling body (6)
6.3. Condensate
pump
The tubing on the peristaltic pump should be replaced annually depending on the
pump duty cycle. Note: This is an important safety item since a pump tube failure
may allow gas to escape since it is at line pressure at this point.
7. Decommissioning
Switch the unit off and drain the condensate, the installation location for the gas
cooler must be frost-free. At temperatures below 0°C, condensate may freeze and
destroy the tubing and filter assemblies.