56
Troubleshooting
12.2 Changing/replacing fuses
Only skilled electricians or service technicians are permitted to replace fuses.
Only replace with fuse types specified by UNION.
12.3 Troubleshooting
no.
Description
1
Cooler/fan does not start up.
•
A check must be made to determine whether the device is properly connected to the
power supply system.
•
The Pury does not activate the calorimeter. The ambient temperature is too high. Switch
the additional readiness condition Diff Ambient to "off" or set a smaller value.
2
Device does not cool.
•
The filter is dirty and must be cleaned.
•
The fan is not delivering enough cooling air. Check fan and or the cooling fins for dirt.
3
Condensate is escaping from the output.
The filter is filled with condensate. Inadequate condensate suction.
•
Check pump hose for free passage.
•
Increase the switching frequency of the condensate pump.
•
The flow rate of the gas is too high. Check whether flow rate exceeds 250 l/h, reduce flow
rate.
Summary of Contents for Pury250 S
Page 1: ...Operating Instructions Gas Cooler Pury250 S ...
Page 5: ...1 Technical data 1 1 Dimensions Weight approx 12 6 kg 28lbs 5 ...
Page 10: ...10 EC Declaration of Conformity ...
Page 15: ...15 4 Protective equipment 4 1 Enclosure cover Fig 4 1 Enclosure cover ...
Page 32: ...32 Transport installation and acceptance ...
Page 34: ...34 Commissioning Switching on ...
Page 36: ...36 Description of the HMI and operator control elements ...
Page 49: ...49 Maintenance ...
Page 53: ...53 ...
Page 54: ...54 ...
Page 58: ...58 Service ...
Page 59: ...59 14 Related documents Declaration of Conformity Service documentation optional ...
Page 60: ...60 Related documents ...
Page 62: ...62 Disposal ...
Page 64: ...64 Spare parts ...