Operation
9.3.4 Difference between ambient and cooler temperatures
Difference is a measure for the degree of drying. Default value is 10°C, setting
range is "Off", 10°C to 30°C.
"Off" means that no temperature difference is permitted. Cooling to the dew point
temperature will be carried out.
9.3.5 Pump condensate quantity
Information on the expected condensate quantity, pump delivers this quantity
automatically at intervals.
Default value is 5 ml/h, setting range is 5 ml/h to 150 ml/h.
Note: The service life of the condensate pump is dependent on the flow rate.
Adjust the condensate quantity to the conditions.
Table 1: Condensate quantity
Condensate quantity for: Dew point 5°C and 250 liter gas saturated at 100 mbar
gauge pressure.
,
0
00
00
,
5
00
,
10
00
,
15
,
00
20
25
00
,
00
,
30
35
00
,
00
,
40
5
10
15
20
25
30
40
45
50
55
60
35
gas temperature
condensate ml/h
Summary of Contents for Pury250 S
Page 1: ...Operating Instructions Gas Cooler Pury250 S ...
Page 5: ...1 Technical data 1 1 Dimensions Weight approx 12 6 kg 28lbs 5 ...
Page 10: ...10 EC Declaration of Conformity ...
Page 15: ...15 4 Protective equipment 4 1 Enclosure cover Fig 4 1 Enclosure cover ...
Page 32: ...32 Transport installation and acceptance ...
Page 34: ...34 Commissioning Switching on ...
Page 36: ...36 Description of the HMI and operator control elements ...
Page 49: ...49 Maintenance ...
Page 53: ...53 ...
Page 54: ...54 ...
Page 58: ...58 Service ...
Page 59: ...59 14 Related documents Declaration of Conformity Service documentation optional ...
Page 60: ...60 Related documents ...
Page 62: ...62 Disposal ...
Page 64: ...64 Spare parts ...