Chapter 8 - Automatic Measurement
UTD2000CEX-II series digital fluorescence oscilloscope can automatically
measure up to 34 parameters. Press the MEASURE button to enter the
automatic measurement menu.
Automatic Measurement menu
Descriptions
Close all Parameters
Pop-up a frame with all
parameters presenting
Open/Close menu for user defined
parameters. When it is on, define
all required parameters with using
the Multipurpose knob.
Real-time dynamically indicates
measurement parameters
Enter advanced measurement menu
Turn off the feature
Automatic calculation of user defined
parameters of average, max, and min.
Only applicable when user defined
parameters are present.
Restart statistics
Clear all measurement
Options
Off
On
On/Off
Off
Parameter 1~5
Off
On
Functions
All Parameters
User Defined
Indicator
Advanced
measurement
Measurement
statistics
Reset statistics
Clear
Advanced measurement menu
Descriptions
Delay measurement
Phase measurement
FRR
: time between the first rising edge of
source 1 and the first rising edge of source 2
FRF
: time between the first rising edge of
source 1 and the first falling edge of source 2
FFR
: time between the first falling edge of
source 1 and the first rising edge of source 2
FFF
: time between the first falling edge of
source 1 and the first falling edge of source 2
LRF
: time between the last rising edge of
source 1 and the last edge of source 2
LRR
: time between the last rising edge of
source 1 and the last rising edge of source 2
LFR
: time between the last falling edge of
source 1 and the last rising edge of source 2
LFF
: time between the last falling edge of
source 1 and the last falling edge of source 2
Options
Functions
Measurement
parameters
Source
edge
UTD2000CEX-II
User Manual